SELF-CHECK PERIOD
2.8
(1)
Set the OFF/PERIOD/FREQUENCY Meter to PERIOD.
(2) Set the OPERATE/SELF-CHECK switch to SELF-CHECK.
(3) Using the four push-buttons on the front panel select, in turn each of the
available numbers of periods over which the measurement is to be made and
check that the following displays are obtained.
TABLE 2
Self-Check (Period) Displays
Number of Periods
Units
Display
1
Sec
0.00001
10
ms
000.010
100
ms
00.0100
1000
ms
0.01000
NOTE:
In both cases check that the decimal point is correctly positioned. If the
above displays are not obtained the instrument should be investigated in
accordance with the procedures in Chapter 6 - iVai ntenance.
2-3
9057/9059