NoiseKen GT-30R Instruction Manual

Disharge gun for electrostatic testing

Advertisement

Quick Links

INSTRUCTION MANUAL
DISCHARGE GUN for ELECTROSTATIC TESTING
Thank you very much for your purchasing this instrument.
This instrument generates high voltage.
Since failure to follow instructions for handling this instrument may
cause an electric shock or other kinds of accidents, please be careful for
safety in handling this instrument.
Please read this booklet carefully, understand its content well to use this
instrument safely and properly, and keep it by your side or other proper
location so that it may be always readily available.
GT-30R
MODEL
GT-30R330
GT-30R2K
GT-30R3302K
NOISE LABORATORY CO., LTD.
Edition 1.07(T)
AEC00264-00E-0H(T)

Advertisement

Table of Contents
loading

Summary of Contents for NoiseKen GT-30R

  • Page 1 INSTRUCTION MANUAL DISCHARGE GUN for ELECTROSTATIC TESTING GT-30R MODEL GT-30R330 GT-30R2K GT-30R3302K • Thank you very much for your purchasing this instrument. • This instrument generates high voltage. Since failure to follow instructions for handling this instrument may cause an electric shock or other kinds of accidents, please be careful for safety in handling this instrument.
  • Page 2 NOTICE • No part of this booklet may be reproduced or transferred, in any form, for any purpose, without the permission of Noise Laboratory Co., Ltd. • The contents of this booklet have been thoroughly checked. However, if a doubtful point, an error in writing or a missing is found, please contact us.
  • Page 3: Important Safety Precautions

    1. IMPORTANT SAFETY PRECAUTIONS The "Important Safety Precautions" explain rules that must be followed to prevent any risk of harm or injury to the user of the instrument or to other people. The instrument may only be used by trained EMC technicians (electrical technicians) There is a risk of death or serious injury, and of the emission of electromagnetic noise that exceeds the stipulated limits.
  • Page 4: Check Package Contents

    For that reason, the product is available in four different sets. This section shows what is supplied with each set. Read it according to the contents of your set. 2–1. GT-30R (Set for Test under IEC 61000-4-2) 2–2. GT-30R2K (Set for 2kΩ Test under ISO 10605) 2–3.
  • Page 5: Gt-30R2K (Set For 2Kω Test Under Iso 10605)

    2. CHECK PACKAGE CONTENTS 2–2. GT-30R2K (Set for 2kΩ Test under ISO 10605) Item Quantity A: Main unit of the discharge gun ·············································· 1 B: 2kΩ testing discharge cup (Attached to the main unit) ··············· 1 C: CR unit 150pF-2kΩ (Attached to the main unit) ························ 1 D: CR unit 330pF-2kΩ...
  • Page 6: Gt-30R330 (Set For 330Ω Test Under Iso 10605)

    2. CHECK PACKAGE CONTENTS 2–3. GT-30R330 (Set for 330Ω Test under ISO 10605) Item Quantity A: Main unit of the discharge gun ·············································· 1 B: 330Ω testing discharge cup (Attached to the main unit) ·············· 1 C: CR unit 150pF-330Ω (Attached to the main unit) ······················ 1 D: CR unit 330pF-330Ω...
  • Page 7: Gt-30R3302K (Set For 330&2Kω Test Under Iso 10605)

    2. CHECK PACKAGE CONTENTS 2–4. GT-30R3302K (Set for 330Ω&2kΩ Test under ISO 10605) Item Quantity A: Main unit of the discharge gun ·············································· 1 B: 330Ω testing discharge cup (Attached to the main unit) ·············· 1 C: 2kΩ testing discharge cup ·················································· 1 D: CR unit 150pF-330Ω...
  • Page 8: Application Form For Instruction Manual

    3. APPLICATION FORM FOR INSTRUCTION MANUAL To: Noise Laboratory Co., Ltd. via sales agent We place an order for an instruction manual. Model Name Serial No. Applicant Address Company Name Department Contact Person Phone No. FAX No. Cut off this page "PURCHASE ORDER FOR INSTRUCTION MANUAL"...
  • Page 10: Table Of Contents

    4. TABLE OF CONTENTS IMPORTANT SAFETY PRECAUTIONS ..................1 CHECK PACKAGE CONTENTS ....................2 2–1. GT-30R (Set for Test under IEC 61000-4-2) ....................2 2–2. GT-30R2K (Set for 2kΩ Test under ISO 10605) .................... 3 2–3. GT-30R330 (Set for 330Ω Test under ISO 10605) ..................4 2–4.
  • Page 11 4. TABLE OF CONTENTS 13. SPECIFICATIONS ........................33 13–1. List of Specifications ..........................33 13–2. Discharge Parameters and Waveform (150pF-330Ω, 330pF-330Ω) ............34 13–3. Discharge Parameters and Waveform (150pF-2000Ω, 330pF-2000Ω) ........... 35 14. CHECKING WAVEFORM ......................36 14–1. Equipment and Configuration ........................36 14–2.
  • Page 12: Preface

    5. PREFACE Thank you very much for your purchase of the discharge gun for electrostatic discharge testing “GT-30R”. It is recommended that the contents of this manual should be thoroughly understood and used as a ready reference for operation. To ensure the safety and procedure of electrostatic tests, thoroughly read this Instruction Manual together with the instruction manual for Electrostatic Discharge Simulator to be connected to this product.
  • Page 13: Basic Safety Precautions For The Safe Use Of The Simulator

    6. BASIC SAFETY PRECAUTIONS FOR THE SAFE USE OF THE SIMULATOR The "Basic Safety Precautions" explain rules that must be followed to prevent damage to property or injury to the user of the instrument or to other people. The symbols below are used to indicate the level of injury or damage that may result if the instrument is used in a way that ignores these precautions.
  • Page 14: Danger Alerts

    6. BASIC SAFETY PRECAUTIONS FOR THE SAFE USE OF THE SIMULATOR 6–2. DANGER Alerts Disassembly Prohibited Do not disassemble or modify Do not remove the cover Failure to comply with the precaution may result in death or serious injury and possible consequences include fire and electric shock For inspection or repair of internal components, please contact your sales agent or the Noise Laboratory repair and calibration center.
  • Page 15 6. BASIC SAFETY PRECAUTIONS FOR THE SAFE USE OF THE SIMULATOR 強制 Safety Rule Only use the instrument with a power supply voltage and frequency that is within the indicated range (AC 100V to 240V, 50Hz/60Hz) Using the instrument with a power supply voltage or frequency outside the indicated range may result in fire or electric shock.
  • Page 16: Caution Alerts

    6. BASIC SAFETY PRECAUTIONS FOR THE SAFE USE OF THE SIMULATOR Do not damage the AC power cable Damage to the AC power cable may result in fire, electric shock, or similar. Take particular care in relation to the following precautions. Do not manipulate the AC power cable Do not bend the AC power cable excessively Do not twist the AC power cable...
  • Page 17 6. BASIC SAFETY PRECAUTIONS FOR THE SAFE USE OF THE SIMULATOR 禁止 Prohibited Do not use the instrument with other than a recommended discharge gun Using the instrument with other than a recommended discharge gun may result in poor operation and abnormal test results.
  • Page 18: Points To Note Regarding Consumables Items

    7. POINTS TO NOTE REGARDING CONSUMABLES ITEMS 7–1. High Voltage Relay This product contains the high-voltage relay. The high-voltage relays are consumable items. The electrical contacts in the high-voltage relays deteriorate with use and this can result in poor electrical connections, contact welding, or insulation failure occurring during normal use.
  • Page 19: Introduction

    8. INTRODUCTION 8–1. How to Read This Document The symbols used in this document and their explanations are shown below. Indicates a supplementary explanation. Indicates where to refer to. Indicates that there is a setting restriction. Indicates that it must be checked before use. 【...
  • Page 20: Terms And Definitions

    8. INTRODUCTION 8–2. Terms and Definitions Term Definition Abbreviation of Electro Magnetic Compatibility. The concept which is generally required on electronic products and devices as a basic ability to radiate emission within the limit and to have enough immunity against external disturbances.
  • Page 21: Name And Function Of Each Part

    External Appearance and Descriptions Figure 9.1 Appearance-1 1 1 1 1 Model number label Indicates Noiseken’s logo, product name and model number. 2 2 2 2 Discharge tips Electrode tips used to perform electrostatic discharges. Use the three types of electrode tips according to what need to be tested.
  • Page 22 For details, see 11–1. Discharge Cup. 4 4 4 4 LED light White LED light used to illuminate EUT. With NoiseKen’s previous simulator models such as ESS-2000AX, the LED light cannot be turned on. Trigger switch Starts electrostatic discharge. The operation varies, depending on the setting of the combined simulator.
  • Page 23 9. NAME AND FUNCTION OF EACH PART Figure9.2 Appearance-2 The serial number label and inspection certification label are placed where the CR unit 8 is VOID seal attached. 13Inspection certification label Serial number label Figure 9.3 Where CR Unit Is Attached...
  • Page 24 If the discharge detection function on the main unit is set to 〔ON〕 , it is turned on in green when a discharge is detected. With NoiseKen’s previous simulator models such as ESS-2000AX, the light is turned on only in red.
  • Page 25: Simulator Main Unit And Connections

    10. SIMULATOR MAIN UNIT AND CONNECTIONS Be sure to turn off the power supply of the main unit when connecting the discharge gun to the main unit or changing setup of equipment. Failure to comply with this rule may cause an electric shock, an injury, or malfunction. Be sure not to insert stuff into the main unit.
  • Page 26: Connection To The Simulator Main Unit

    Figure 10.1 High-voltage Connector ESS-S3011 GT-30R GT-30R ESS-B3011 Connection to ESS-B3011 Connection to ESS-S3011 Figure 10.2 Connection to Main Unit This product can also be connected to NoiseKen’s previous simulator models such as ESS-2000AX. For details, see 5–2. Combinations with Simulators.
  • Page 27: How To Replace Component Units

    11. HOW TO REPLACE COMPONENT UNITS 11–1. Discharge Cup When demounting or mounting the discharge cup, turn off the electrostatic discharge simulator or make it 【 【 【 【 STOP】 】 】 】 status. Failure to follow this instruction might cause an electric shock or an injury. Use the discharge cup with the correct CR unit.
  • Page 28: Replacing Discharge Cup

    11. how to replace component units Replacing Discharge Cup ID label Attach the 330Ω discharge cup discharge tip 2kΩ discharge cup Attach to the main unit of the discharge gun Figure 11.1 Appearance of Discharge Cup To remove the discharge cup, turn it slowly anticlockwise while holding it firmly by hand.To attach it, turn it clockwise facing the main unit of the gun, until it locks, making a click sound.
  • Page 29: Discharge Tip

    11. how to replace component units 11-2. Discharge Tip Before replacing the discharge tip, set the simulator to the stop state 〔 〔 〔 〔 STOP〕 〕 〕 〕 Failure to comply with this rule may result in electric shock or injury. Before touching the discharge tip, remove static charge After an electrostatic test, the discharge tip may still be charged at a high voltage.
  • Page 30: Replacing Discharge Tip

    11. how to replace component units Replacing Discharge Tip After removing static charge from the discharge tip, remove the tip by turning it anticlockwise while holding the discharge cup so that the cup does not become loose. Turn the desired discharge tip clockwise to attach it to the discharge gun firmly. Figure 11.3 Replacing Discharge Tip...
  • Page 31: Cr Unit

    11. how to replace component units 11-3. CR Unit Before replacing the CR unit, set the simulator to the〔 〔 〔 〔 STOP〕 〕 〕 〕 state Failure to comply with this rule may cause an electric shock, an injury, or malfunction. Before removing the CR unit, remove static charge After an electrostatic test, the CR unit may still be charged at a high voltage.
  • Page 32: Gnd Clip

    11. how to replace component units 11-4. GND Clip The alligator clip at the end of the GND cable can be detached. When grounding the GND cable with a screw, remove the clip as shown below. GND clip cover M3 round terminal Alligator clip Figure 11.5 How to Detach GND Clip...
  • Page 33: Operation

    12. OPERATION This Operation Manual deals with the operation of Discharge Gun alone. For operating procedures for the entire testing system, refer to the operation manual for the test system to be used in combination. 12–1. Starting Test Sufficient care must be taken during the test, as the voltage set in the simulator is generated.
  • Page 34: Led Light

    During the test, the light can be used for irradiation by enabling the LED light function 〔ON〕. When the main unit of the simulator is started, the LED light is turned on. Figure 12.2 LED Light With NoiseKen’s previous simulator models such as ESS-2000AX, the LED light cannot be turned on. 12–3. Stopping a Test Temporary stop by the trigger switch does not set the 〔...
  • Page 35: Specifications

    13. SPECIFICATIONS 13–1. List of Specifications Parameters Function/Performance The compliance standard varies, depending on the discharge cup and CR unit. Discharge cup CR unit Compliance standard IEC 61000-4-2 Ed1.2 150pF-330Ω IEC 61000-4-2 Ed2.0 Discharge cup for 330Ω tests ISO 10605 2 Compliance standard 330pF-330Ω...
  • Page 36: Discharge Parameters And Waveform (150Pf-330Ω, 330Pf-330Ω)

    13. SPECIFICATIONS 13–2. Discharge Parameters and Waveform (150pF-330Ω, 330pF-330Ω) Current value I Current value I Setup ±30% ±30% Peak current [A] Rise time voltage (150pF: at 30ns) (150pF: at 60ns) ±10% [ns] [kV] (330pF: at 65ns) (330pF: at 130ns) 0.7 to 1.0 15.0 0.7 to 1.0 22.5...
  • Page 37: Discharge Parameters And Waveform (150Pf-2000Ω, 330Pf-2000Ω)

    13. SPECIFICATIONS 13–3. Discharge Parameters and Waveform (150pF-2000Ω, 330pF-2000Ω) Current value I Current value I Setup ±30% ±30% Peak current [A] Rise time voltage (150pF: at 180ns) (150pF: at 360ns) -0%, +30% [ns] [kV] (330pF: at 400ns) (330pF: at 800ns) 0.7 to 1.0 0.55 0.30...
  • Page 38: Checking Waveform

    14. CHECKING WAVEFORM 14–1. Equipment and Configuration For observing the output waveform of this product, equipment shown in Figure 14.1 below, such as the Faraday cage, target, and oscilloscope (the frequency range: 2GHz or more for IEC 61000-4-2, and 1GHz or more for ISO 10605) is required Discharge gun Current target GND cable of the...
  • Page 39: Related Accessories

    15. RELATED ACCESSORIES 330Ω discharge cup: 03-00071A 2kΩdischarge cup: 03-00072A For IEC 61000-4-2 and ISO 10605 330Ω tests For ISO 10605 2kΩ tests CR unit 150pF-330Ω: 06-00073B CR unit 330pF-330Ω: 06-00075B For IEC 61000-4-2 and ISO 10605 330Ω tests For ISO 10605 330Ω tests CR unit 150pF-2kΩ: 06-00074B CR unit 330pF-2kΩ: 06-00076B For ISO 10605 2kΩ...
  • Page 40: Calibration

    16. CALIBRATION Both of the ESD immunity tests conforming to the IEC standard and the ISO standard are available on this unit with changing the gun head and/or the CR unit of the discharge gun. In case of calibrating this unit, specify the combination of 1) the main body of the discharge gun 2) the gun head 3) the CR unit.
  • Page 41: Warranty

    17. WARRANTY Servicing terms The following terms are applicable to servicing by Noise Laboratory Co., Ltd., (hereafter referred to as the Company) provided to maintain the intended performance of its products. Scope The following terms shall apply only to products made by the Company. Technical servicing fee In the event of a failure of a product within the warranty period (see warranty section), the Company will repair a product without charge.
  • Page 42 17. WARRANTY Limited Warranty Noise Laboratory Co., Ltd. (hereafter referred to as the Company) warrants its products to be free from defects in materials and workmanship under normal use and service for a period of one year from date of delivery. In the event of failure of a product covered by this warranty, the Company will repair the product or may, at its option, replace it in lieu of repair without charge.
  • Page 43: Maintenance

    18. MAINTENANCE When repair, maintenance or internal adjustment of the unit is required, a qualified service engineer takes charge of such work. Maintenance on the user side is restricted to the outside cleaning and functional check of the unit. When checking or replacing the fuse, turn off the switch of the unit and disconnect the plug socket beforehand.
  • Page 44: Contacting Technical Support

    When it is necessary to send your unit back to Noise Laboratory, fill in the repair order form completely, pack the unit in the original package or equivalent one suitable for transit, and send the package. Repair and Calibration Center TEL +81-42-712-2051 FAX +81-42-712-2050 E-mail:sales@noiseken.com...
  • Page 46 NOISE LABORATORY CO., LTD. Produced by: Noise Laboratory Co., Ltd. 1-4-4, Chiyoda, Chuo-ku, Sagamihara City, Kanagawa Pref., 252-0237, Japan TEL: +81-(0)42-712-2031 FAX: +81-(0)42-712-2030 URL: http://www.noiseken.co.jp Printed in Japan...