Parts that are broken, missing, plainly worn, distorted or • contaminated should be replaced immediately . Should a repair or replacement become necessary, Dispomed • recommends placing a request by phone or in writing to the nearest Dispomed regional service center .
Page 4
TM/MC ANESTHESIA MACHINE The Moduflex Coaxial is the first machine to incorporate a CO2 absorber with an integrated Coaxial breathing circuit . This reduces clutter, eliminates the possibility of a disconnection, and reduces heat loss by preheating the inhaled gasses with the exhaled gasses . The Moduflex Elite is the same as the Moduflex Coaxial, but with the traditional “Y”...
RECEIVING AND ASSEMBLY TM/MC INSTRUCTIONS Coaxial Elite Your Moduflex Coaxial or Elite anesthesia ma- chine requires some assembly before use . The Moduflex Elite is the same as the Moduflex Coaxial, but with the traditional “Y” connec- tions . RECEIVING...
Page 6
. cur) and slide the U-channel onto the post of the base assembly (the absorber connections face the same way as the locking castors) . DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-025...
Page 7
RECEIVING, ASSEMBLY TM/MC INSTRUCTIONS CONTINUED Remove the accessories from the bag . Insert and tighten the four levelling screws at the top and the bottom of the post . 805-0975-025...
Page 8
When connecting a Universal F circuit (not included) pay attention to the inspiration and expiration outlets to hook them up correctly . Expiration Universal F Breathing Circuit SKU: 940-12U6006 Inspiration DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-025...
Page 9
RECEIVING, ASSEMBLY TM/MC INSTRUCTIONS CONTINUED Connect the fresh gas hose from the fresh gas Install the breathing bag on the swivel bag outlet to the CO2 fresh gas inlet . mount . CAUTION: Always depress locking mechanism while inserting and removing fresh gas connec- tion .
Page 10
Fill the canister with absorbent using loose-fill granules . (Caution: do not overfill the canister; leave approximately ¼” (5 mm) from the top empty.) Reinstall the canister and tighten the sealing screw . DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-025...
. The Bain adap- tor is equipped with a momentary valve to avoid forgetting the valve in a closed position, which could cause pulmonary trauma to the animal . DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-025...
Page 13
OPTIONAL ACCESSORIES TM/MC ASSEMBLY INSTRUCTIONS SCAVENGER INTERFACE CIRCUIT ALARM Screw the interface to anesthesia unit using the Screw the circuit alarm holder on the anes- screws provided in the package . Install the scav- thesia machine . Screw the circuit alarm to the enger interface as per interface instructions .
Obstruct the patient’s circuit outlet with a plug gas outlet of the anesthesia machine . or your thumb . Open flowmeter to approximately 200 ml . Close the gas evacuation valve . DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-025...
Page 17
DAILY LEAK TM/MC CHECKS PROCEDURES Press oxygen flush button and hold until the Check the manometer . The pressure must not pressure on the manometer indicates approxi- drop (but it can increase) . If the pressure drops, mately 30 cm H2O . replace the patient circuit and the bag and try again .
ANNUAL TM/MC MAINTENANCE 845-S113 845-S106 Prerequisite: Preventative maintenance kit # 975-0900-900 The maintenance kit contains all the necessary parts to per- form the annual maintenance . Refer to this technical draw- ing the replacement parts are 980-0200-055 identified . Perform the daily leak check procedure (page 845-S010 17) once after the mainte-...
. to such damage, injury or loss, the product SERVICE AND INFORMATION CENTER was: Dispomed Ltd is proud to offer its customers Damaged, misused or misapplied; • an online or telephone service to answer Repaired, altered or modified by a any questions or concerns regarding its •...
Page 22
TM/MC Manufactured by Dispomed ltd Thank you for purchasing a Dispomed product. If you have any questions, please feel free to contact customer service . 745 Nazaire-Laurin Joliette (QC) Canada J6E 0L6 T . 1-800-363-1746 (Canada and U .S .A only) T .