Dispomed regional service center . • This product, or any of its parts, should not be repaired other than in accordance with written instructions provided by Dispomed and by Dispomed trained personnel . • The product may not be altered without prior written approval from Dispomed .
RECEIVING AND ASSEMBLY TM/MC INSTRUCTIONS Your Moduflex Access2 anesthesia machine requires some assembly . RECEIVING In order to reduce the possibility of damage to a minimum during shipping, certain components have been detached from the unit . These components are packed separately within the shipping crate .
Page 6
RECEIVING, ASSEMBLY TM/MC INSTRUCTIONS CONTINUED Remove the accessories found in the bag Remove the wheeled base and post from the inside the large box . box . DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-650...
Page 7
RECEIVING, ASSEMBLY TM/MC INSTRUCTIONS CONTINUED Slide the tapered end of the black post into Remove the head assembly from the inner the base . box and slide it onto the top of the black post . 805-0975-650...
Page 8
Install the vaporizer as indicated on page head using the Allen key supplied . 12 . Install a breathing bag on the rotary bag mount of the absorber . DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-650...
Page 9
RECEIVING, ASSEMBLY TM/MC INSTRUCTIONS CONTINUED Fill the canister with absorbant using Install the patient circuit on the connections pre-packed or loose-fill granules on the front of the absorber . (caution: do not overfill the canister; leave approximaly 1/4” (5mm) from the top empty) . Install the canister and tighten the sealing screw .
Page 10
2 wrenches (5/8 and 11/16) hose to a 50 p .s .i . (3 bar) oxygen supply . in order to reduce the risk of accidentally disconnecting the hose . DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-650...
. IMPORTANT : Always refer to the user manual of the option you are installing for more detailed instructions . DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-650...
Page 13
OPTIONAL ACCESSORIES TM/MC ASSEMBLY INSTRUCTIONS MODUFLEX BAIN SCAVENGER INTERFACE Install Bain circuit adaptor as illustrated . Install the scavenger interface as per interface instructions . A Bain type open circuit on an anesthesia machine gives you better control during surgery . The During surgery the unused gas should be evacuated breathing bag is always inflated, no need to wait for outside of the building to avoid it being inhaled...
Page 14
. IMPORTANT : Always refer to the user manual of the option you are installing for more detailed instructions . DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-650...
Page 15
OPTIONAL ACCESSORIES TM/MC ASSEMBLY INSTRUCTIONS CIRCUIT ALARM WALL-MOUNT Installation of the patient circuit pressure Install wall-mount adaptor as illustrated . alarm . During assisted ventilation, the alarm will alert you if the predetermined pressure level is reached . For example, if you forget the evacuation valve in a closed position, the circuit alarm will alert you with a warning sound at the predetermined pressure .
DAILY LEAK TM/MC CHECK PROCEDURE The unit’s integrity should be verified every day, and at each Prerequisites: replacement of patient circuit components such as breathing Oxygen supply of 40 to 60 bags, tubes etc . A properly maintained anesthesia machine psi (3 bar) Breathing circuit reduces gas consumption and operating room pollution .
Page 18
At the end of each day, it is recommended that you close the oxygen tank and fill the anesthetic agent vaporizer . Caution: Open the evacuation valve before using the anesthesia machine again . DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-650...
ANNUAL TM/MC MAINTENANCE Prerequisite: Preventative maintenance kit # 975-0900-900 The maintenance kit 980-0300-180 contains all the necessary 845-S011 parts to perform the annual maintenance . Refer to this technical drawing the parts to be replaced are identified . It is important to perform the daily leak check procedure (page 17) once the maintenance has been...
. (5) years from the date of delivery to the Dispomed Ltd . is not responsible for deterioration, wear or abuse of this unit . customer, provided that it is used according...