Summary of Contents for dispomed Moduflex Access 2

  • Page 1 Printed in Canada...
  • Page 2: Table Of Contents

    Order Part Numbers . . . . . . . . . . . . . . . . . . 22 - 23 DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-650...
  • Page 3: User Responsibility

    Dispomed regional service center . • This product, or any of its parts, should not be repaired other than in accordance with written instructions provided by Dispomed and by Dispomed trained personnel . • The product may not be altered without prior written approval from Dispomed .
  • Page 4: Standard And Optional Features

    Patient circuit with 2 and 3 liter bags and fresh gas oulet • 5 year warranty • OPTIONAL FEATURES Dual cylinder holder • Bain circuit adaptor • Circuit pressure alarm • Shelf assembly • MRI Compatibility • DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-650...
  • Page 5: Receiving And Assembly Instructions

    RECEIVING AND ASSEMBLY TM/MC INSTRUCTIONS Your Moduflex Access2 anesthesia machine requires some assembly . RECEIVING In order to reduce the possibility of damage to a minimum during shipping, certain components have been detached from the unit . These components are packed separately within the shipping crate .
  • Page 6 RECEIVING, ASSEMBLY TM/MC INSTRUCTIONS CONTINUED Remove the accessories found in the bag Remove the wheeled base and post from the inside the large box . box . DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-650...
  • Page 7 RECEIVING, ASSEMBLY TM/MC INSTRUCTIONS CONTINUED Slide the tapered end of the black post into Remove the head assembly from the inner the base . box and slide it onto the top of the black post . 805-0975-650...
  • Page 8 Install the vaporizer as indicated on page head using the Allen key supplied . 12 . Install a breathing bag on the rotary bag mount of the absorber . DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-650...
  • Page 9 RECEIVING, ASSEMBLY TM/MC INSTRUCTIONS CONTINUED Fill the canister with absorbant using Install the patient circuit on the connections pre-packed or loose-fill granules on the front of the absorber . (caution: do not overfill the canister; leave approximaly 1/4” (5mm) from the top empty) . Install the canister and tighten the sealing screw .
  • Page 10 2 wrenches (5/8 and 11/16) hose to a 50 p .s .i . (3 bar) oxygen supply . in order to reduce the risk of accidentally disconnecting the hose . DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-650...
  • Page 11 ASSEMBLED TM/MC VIEWS 1 805-0975-650...
  • Page 12: Optional Accessories Assembly Instructions

    . IMPORTANT : Always refer to the user manual of the option you are installing for more detailed instructions . DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-650...
  • Page 13 OPTIONAL ACCESSORIES TM/MC ASSEMBLY INSTRUCTIONS MODUFLEX BAIN SCAVENGER INTERFACE Install Bain circuit adaptor as illustrated . Install the scavenger interface as per interface instructions . A Bain type open circuit on an anesthesia machine gives you better control during surgery . The During surgery the unused gas should be evacuated breathing bag is always inflated, no need to wait for outside of the building to avoid it being inhaled...
  • Page 14 . IMPORTANT : Always refer to the user manual of the option you are installing for more detailed instructions . DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-650...
  • Page 15 OPTIONAL ACCESSORIES TM/MC ASSEMBLY INSTRUCTIONS CIRCUIT ALARM WALL-MOUNT Installation of the patient circuit pressure Install wall-mount adaptor as illustrated . alarm . During assisted ventilation, the alarm will alert you if the predetermined pressure level is reached . For example, if you forget the evacuation valve in a closed position, the circuit alarm will alert you with a warning sound at the predetermined pressure .
  • Page 16 ASSEMBLED TM/MC VIEWS 2 DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-650...
  • Page 17: Daily Leak Check Procedure

    DAILY LEAK TM/MC CHECK PROCEDURE The unit’s integrity should be verified every day, and at each Prerequisites: replacement of patient circuit components such as breathing Oxygen supply of 40 to 60 bags, tubes etc . A properly maintained anesthesia machine psi (3 bar) Breathing circuit reduces gas consumption and operating room pollution .
  • Page 18 At the end of each day, it is recommended that you close the oxygen tank and fill the anesthetic agent vaporizer . Caution: Open the evacuation valve before using the anesthesia machine again . DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-650...
  • Page 19: Annual Maintenance

    ANNUAL TM/MC MAINTENANCE Prerequisite: Preventative maintenance kit # 975-0900-900 The maintenance kit 980-0300-180 contains all the necessary 845-S011 parts to perform the annual maintenance . Refer to this technical drawing the parts to be replaced are identified . It is important to perform the daily leak check procedure (page 17) once the maintenance has been...
  • Page 20: Consumables Materials List

    # 940-321525 # 905-8855-000 1 liter: 18.9 liters: # 940-321625 # 905-8862-000 2 liters: # 940-321725 3 liters: # 940-321825 GAS EVACUATION TUBING KRYTOX 5’: # 850-0052-000 # 985-0060-000 DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-650...
  • Page 21: Warranty

    . (5) years from the date of delivery to the Dispomed Ltd . is not responsible for deterioration, wear or abuse of this unit . customer, provided that it is used according...
  • Page 22: Order Part Numbers

    ORDER TM/MC PART NUMBERS Gas evacuation valve Flowmeter shield 975-0650-500 930-0100-110 Dome, compact 980-0300-155 flowmeter valve -20 +100 CM gauge 930-0105-000 900-6050-000 Preassembled flush cartridge 975-0300-275 Sodalime canister 980-0300-080 DISPOMED Ltd. / 1 800 363-1746 / info@dispomed.com / www.dispomed.com 805-0975-650...
  • Page 23 REORDER TM/MC PART NUMBERS Fresh gas elbow 850-0030-000 Quick Connect 850-LC10004 805-0975-650...
  • Page 24 MODUFLEX ACCESS by DISPOMED TM/MC Manufactured by Dispomed ltd Thanks for purchasing a Dispomed product. For further assistance, contact customer support. 745 Nazaire-Laurin Joliette (QC) Canada J6E 0L6 Phone: 1-800-363-1746 (Canada et États-Unis seulement) Phone: 1-450-759-9395 Fax: 1-450-759-8181 info@dispomed.com www.dispomed.com...