Clean The Cusa Excel Console; Clean The Sterilizer Case; Maintain The Handpiece - Valleylab CUSA EXcel-8 Service Manual

Table of Contents

Advertisement

Clean the CUSA EXcel Console
Frequency
Equipment
Task
Once a year
Tip torquing wrench
Check the torque.
Once a year
CUSA console
Contact a Valleyiab service
representative to replace the cooling
water pump tubing.
Clean the CUSA EXcel Console
You cannot sterilize the CUSA console, but you can clean it, disinfect it, or
both. Use a mild cleaning solution or disinfectant and a damp cloth.
Electric Shock Hazard
Always unplug the CUSA system before cleaning.
Notice
Do not rub, press, or touch any panets with solvents; caustic, corrosive, or
abrasive cleaning or disinfectant compounds; or other materials that could
scratch the panels. Do not use a betadine-based solution; it will cause
discoloration.
Do not allow fluids to enter the chassis.
1.
Unplug the unit power cord from the wall receptacle.
2.
Using standard procedures for your facility, thoroughly wipe all
surfaces and cords with a cleaning solution and disinfectant.
3.
Make sure that the surface is completely dry before using the system
again.
Clean the Sterilizer Case
Use a neutral detergent to clean the sterilizer case.
Notice
Do not clean the sterilizer case with abrasives. Produci damage will result.
Maintain the Handpiece
Maintaining the handpiece involves cleaning the handpiece, lubricating
the handpiece connector o-rings with a high viscosity lubricant such as
silicone grease, or replacing the handpiece connector o-rings. If a problem
develops, you may also need to have the handpiece recalibrated.
CUSA EXcel-8 System Service Manual
9-3