Invacare® M6, M9, D Cylinder with Continuous Flow Regulator For HomeFill® Oxygen System en IOH2M6, IOH2M9, IOH2MD User Manual This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference.
Page 2
All rights reserved. Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. Trademarks are identified by ™and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.
Invacare® M6, M9, D Cylinder with Continuous Flow Regulator 1 General Read the manual 1.1 Symbols No open flame Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage.
General 1.2 Intended Use intended to be used by or under adult supervision only after reading and understanding the instructions and warnings of this user manual. This cylinder assembly is intended to store, regulate, and deliver supplemental oxygen to patients who may have difficulty extracting Only personnel instructed and trained in the use of the HomeFill®...
Do not return products to our factory without from Invacare or a dealer, with a copy of the seller’s invoice required prior consent. C.O.D. shipments will be refused; please prepay for coverage under this warranty.
Page 7
THE TERMS CONTAINED HEREIN. THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN. INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER. SOME STATES DO NOT ALLOW THE EXCLUSION OR...
– If you are unable to understand the warnings, cautions professional users. or instructions, contact a healthcare professional, – This cylinder is to be used with the Invacare HomeFill dealer or technical personnel before attempting to oxygen system. This cylinder contains 93±3% oxygen use this equipment.
Page 11
If you use these types of cigarettes or other sources of ignition out of the room products Invacare suggests you wash your hands prior in which this cylinder is located and away from where to using the HomeFill system. If the cylinder fillport oxygen is being delivered.
Page 12
Invacare® M6, M9, D Cylinder with Continuous Flow Regulator DANGER! WARNING! Risk of Fire Risk of Injury or Death Dirt or contaminants on the cylinder assembly or Contaminants in the oxygen may cause injury or death. regulator can cause fire, injury or death.
Page 13
To avoid injury or accessories may vary by manufacturer. Invacare accessories death: such as the cylinder bag and Invacare oxygen tubing are latex – The cannula MUST be routed and secured properly. and phthalate free. Refer to the packaging on any cylinder –...
Page 14
Invacare® M6, M9, D Cylinder with Continuous Flow Regulator The cylinder assembly MUST be emptied of gas prior to shipping by courier or postal service. Refer to Emptying the Cylinder Assembly. Because the cylinder contains magnetic, ferrous material, it is not suitable to be in an MRI room.
Refer to The Hydrostatic use only. Not to be filled for resale or use by Testing Date section. professional users. – Filling of cylinder assemblies is performed by Invacare WARNING! HomeFill Oxygen System only. Risk of Injury –...
Invacare® M6, M9, D Cylinder with Continuous Flow Regulator DANGER! WARNING! Risk of Injury or Death Risk of Injury Improper filling or abuse of this cylinder assembly, or To reduce the risk of injury: failure to follow these safety instructions may cause –...
– All cylinders MUST be inspected before attempting to fill. – If the cylinder is found to be defective or does not work properly, contact an Invacare dealer or qualified technician for service or replacement. Before filling a cylinder assembly, read and follow the information in Prior to each use, inspect product for visible damage.
Page 18
Invacare® M6, M9, D Cylinder with Continuous Flow Regulator Position the cylinder assembly A so the outlet connection D is pointing away from the user and any other person(s). WARNING! Risk of Injury To reduce the risk of injury: – DO NOT direct flow of oxygen at any person, or flammable material when adjusting the flow dial E.
The regulator should be inspected and tested at a minimum every 5 years. This inspection and test MUST be performed by Invacare DANGER! Corporation or by an Invacare approved repair center. Risk of Fire...
Invacare® M6, M9, D Cylinder with Continuous Flow Regulator Follow the cleaning/disinfectant product manufacturer’s instructions CAUTION! for use, handling, storage and disposal of the product. Risk of Damage – DO NOT completely empty the cylinder. Otherwise, Follow this procedure to clean the cylinder assembly as required damage to the cylinder could occur.
WARNING! Risk of Injury or Damage Before transferring the product to a different user, Use of non-Invacare accessories may result in serious providers MUST inspect the product for damage and injury or damage. – Invacare products are specifically designed and disinfect the product.
6.2 Wear and Tear Invacare reserves the right to ask for any item back that has an alleged defect in workmanship. See Warranty that shipped with the product for specific warranty information.
Troubleshooting 7 Troubleshooting 7.1 Troubleshooting If the Regulating Cylinder fails to function, consult the Troubleshooting guide. If problem cannot be corrected, consult your Dealer/Home Care equipment provider. Problem Possible Cause Solution No flow. Flow dial at setting “0”. Adjust flow dial to prescribed setting. Cylinder empty.
The cylinder assembly is filled with USP 93±3% oxygen using its proprietary fill port that allows filling only through the use of an Invacare HomeFill System. The duration of oxygen flow between fills is dependent on the size of the cylinder and the flow setting.
Page 25
Storage Temperature Range: -40°F to 140°F (–40°C to 60°C) Max Storage Humidity: 95% RH Assembly Dimensions Model: IOH2M9 IOH2MD IOH2M6 Diameter: 4.38 in (11.13 cm) 4.38 in (11.13 cm) 3.0 in (7.62 cm) Weight: 4.8 lb (2.18 kg) 7 lb (3.18 kg) 3.5 lb (1.60 kg)
Page 26
Invacare® M6, M9, D Cylinder with Continuous Flow Regulator Cylinders Cylinder Size: Volume (H 0.871 L (0.031 ft 1.46 L (0.052 ft 2.48 L (.088 ft Oxygen Capacity: 148 L (5.23 ft 248 L (8.76 ft 422 L (14.9 ft...