To Install the Detection Module

P
eering Gas Sensing Technology.
ion
Ensure the pneumatic tubes are
properly identified for later connection
to the correct connectors.
Disconnect the tubes only from the
Detection Module, not from the
Housing Module. The tubes must
remain connected at all times to the
barbs on the interior face of the flame-
arrestors.
For complete removal of the Detection
Module: a half-turn, counter-clockwise,
of the Luer connectors will remove the
pneumatic tubes:
two on the left
one on the right.
You can now remove the Detection
Module completely.
Check the identification of the pneumatic tubes and connectors.
Connect the tubes to the connectors: a half-turn, clockwise, of the
Luer connectors will secure the pneumatic tubes:
two on the left
one on the right.
Port 3 remains open.
Ensure the tubes have no twists or
kinks.
Hold the ribbon cable out of your way while you insert the
Detection Module.
There are locating pins to ensure correct positioning of the
Detection Module on the base-plate.
Move the module from side to side until it locks into place.
Ensure the tubes have no twists or kinks.
Page 31 of 66
TITAN 2 Instrument User Manual V1.0
ionscience.com