ION TITAN 2 User Manual page 19

Table of Contents

Advertisement

80
24
100
24
1000
16
1100
32
1200
36
P
eering Gas Sensing Technology.
ion
Returns block of data indicating the state of the sensor
temperature controller. Six 32-bit integers are returned.
e.g.
00000000 84 13 00 00 22 02 00 00 AD 08 00 00 A0 0F 00 00 „···"···­··· ···
00000010 01 00 00 00 01 00 00 00
These can be decoded as follows:
84 12 00 00 = 0x00001284 = 4740
The values have the following meaning
Integer
Meaning
1
Temperature in degrees Celsius x 100
2
Ambient temperature in degC x 16
3
Heater drive power in % x 100
4
Heater driver voltage
5
1 if required temperature has been reached
6
Time in seconds taken to reach target from
switch on
Returns block of data indicating state of background
temperature controller. Six 32-bit integers are returned.
e.g.
00000000 FC 44 00 00 21 02 00 00 00 00 00 00 01 00 00 00 üD··!···········
00000010 EF 00 00 00 02 00 00 00
Integer Meaning
1
Heater Temperature in degC x 100
2
Ambient temperature in degC x 16
3
Heater drive power in % x 100
4
Heater driver voltage
5
Power supply level x 10
6
Heater level
Unique ID of Titan 2 – a string of 16 ASCII characters
unique to each Titan 2
Firmware versions of Titan 2 and Flow Controller. Two null
terminated strings are returned. e.g.
00000000 56 30 2E 31 2E 33 33 00 56 30 2E 31 2E 31 31 00
V0.1.33·V0.1.11·
00000010 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 ················
Returns a block of data indicating the state of the flow.
This block contains 9 32-bit integers. e.g.
00000000 21 03 00 00 DA 0D 00 00 20 2F 00 00 88 00 00 00 !···Ú···
/··ˆ···
00000010 01 00 00 00 00 00 00 00
These can be decoded as follows:
21 03 00 00 = 0x00000321 = 801
The values have the following meaning:
Integer Meaning
1
Current flow x 100
2
Flow Differential pressure sensor reading
3
Absolute pressure sensor reading
4
Pump power output
TITAN 2 Instrument User Manual V1.0
········
ï·······
········
Page 19 of 66
ionscience.com

Advertisement

Table of Contents
loading

Table of Contents