Introduction To Titan 2 - ION TITAN 2 User Manual

Table of Contents

Advertisement

Introduction to TITAN 2

The TITAN 2 is a fixed benzene monitor certified for Zone 1 hazardous areas. It detects 0.02 - 20 ppm concentrations in
ambient air, taking one sample per minute.
It monitors both the current benzene concentration and the STEL (Short-Term Exposure Limit, calculated over the
previous 15 minutes).
Real-time display of the measurement is on the LCD and transmitted on a 4-20 mA channel. Historic data stored in Titan 2
can be viewed and stored externally either over the RS485 or the USB connection.
Alarms and warnings are transmitted by the 4-20 mA channel and by two relays programmable to be Normally Open or
Normally Closed.
It is powered by a single power supply in the range of range of 19VDC @ 4A to 32VDC @ 2.4A.
The Titan 2 has two modules:
The Housing Module that is
installed permanently.
The Detection Module that can be
removed for servicing and
calibration.
All the operational and calibration data
is stored within the Detection Module.
After the Detection Module has been
calibrated, it will function correctly
when installed in any Housing Module.
A Detection Module can be calibrated
and tested off-site, then taken to the
installation site and installed in place of
an existing module. The removed
module can then be serviced ready for
installation later or in another Housing
Module.