ION TITAN 2 User Manual page 30

Table of Contents

Advertisement

P
eering Gas Sensing Technology.
ion
Two captive, hex-socket screws hold the
Detection Module in the Housing
Module, as shown.
Two holes in the front of the Detection
Module give access to the screws.
Use a 3mm allen-key to reach the
screws directly behind the holes. Red
circle indicates entry,
Turn each screw counter-clockwise until
it is completely loose.
When the captive screws are
completely loose, grip the sides of
the frame of the Detection Module,
as shown, to remove it from the
Housing Module.
Be careful not to cause damage the
PCBs.
Be careful to remove it only far
enough to disconnect the Luer
connectors of the 3 pneumatic
tubes.
Page 30 of 66
TITAN 2 Instrument User Manual V1.0
ionscience.com