ION TITAN 2 User Manual page 21

Table of Contents

Advertisement

1300
1400
1500
4
1600
4
1604
4
1608
4
1700
28
1800
F000
44
F080
16
F100
128
P
eering Gas Sensing Technology.
ion
This is decoded as 8 32-bit numbers. The last two integers
are not used when writing.
Integer
Meaning
1
Year
2
Month
3
Day
4
Hour
5
Minute
6
Second
7
Day of week
8
Day of year
Used during code upgrade. For ISL use only. Reading or
Writing to this address may require you to send the Titan 2
to a Service Center.
Used during upgrade of flow controller. For ISL use only.
Reading or Writing to this address may require you to send
the Titan 2 to a Service Center.
Override control of measuring sequence. For ISL use only.
Override control of pump. For ISL use only.
Override control of by-pass pump. For ISL use only.
Override control of sample valve. For ISL use only.
Read latest result data.
A block of 7 32-bit integers is returned
Integer
Meaning
1
Benzene value in ppm x 10
2
Results Flags
3
Flag indicating start of measurement cycle
4
Diagnostic Step in measurement cycle
5
Last sensor reading
6
Stel result in ppm x 10
7
Last fault
Writing to any space in this address range will clear flag
indicating start of measurement cycle (integer 3).
Diagnostic, test control of 4-20mA output. For Ion Science
use only.
Block of data controlling file transfer to/from internal
store. Write only.
Block of data giving information on file transfer to/from
internal store. Read only.
Used to read/write data from/to files on internal store.
TITAN 2 Instrument User Manual V1.0
Page 21 of 66
ionscience.com