ION TITAN 2 User Manual page 64

Table of Contents

Advertisement

No signal from PID Light
sensor.
Memory full:
Not enough memory
space to store history.
PID ADC Fail:
Communication failure
Step overrun:
Internal hardware fault
Step error:
Software fault.
Watchdog:
Software fault.
Pump fault:
Failed communication
with pump board.
Pump power fail:
Pump power fault.
P
eering Gas Sensing Technology.
ion
PID Lamp not illuminated.
Sensor failure.
During Titan 2 start-up,
wait at least 5 cycles for
the lamp to illuminate.
Memory full.
Use Titan PC to download
or delete History Record
Files.
Internal hardware fault.
Remove the Detector
Module to check that the
black wiring loom is
connected and not
damaged.
Program code step took
more than 0.5s to execute.
Contact the Service Centre.
Invalid sequence step.
Contact the Service Centre.
Software fault.
Contact the Service Centre.
Internal hardware fault.
Contact the Service Centre.
Pump board power failure.
Contact the Service Centre.
Page 64 of 66
TITAN 2 Instrument User Manual V1.0