Testboy TV 465 Instruction Manual
Hide thumbs Also See for TV 465:
Table of Contents

Advertisement

Quick Links

®
Testboy
TV 465
Instruction manual
Version 1.0

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TV 465 and is the answer not in the manual?

Questions and answers

Summary of Contents for Testboy TV 465

  • Page 1 ® Testboy TV 465 Instruction manual Version 1.0...
  • Page 2 Mark on your equipment certifies that this equipment meets the requirements of the EU (European Union) concerning safety and electromagnetic compatibility regulations © 2012 TESTBOY No part of this publication may be reproduced or utilized in any form or by any means without permission in writing from TESTBOY.
  • Page 3: Table Of Contents

    ® Metrel d.d. TV 465 Table of contents Table of contents General description....................5 Warnings .......................6 Battery and charging .....................7 New battery cells or cells unused for a longer period ..........8 Standards applied....................9 Instrument description ..................10 Front panel ......................10 Connector panel ....................11 Back side ......................12...
  • Page 4 ® Metrel d.d. TV 465 Table of contents 5.2.6 Substitute leakage - P ..................34 5.2.7 Polarity test....................36 5.2.8 Functional test ....................37 5.2.9 Voltage TRMS ....................38 Autotest sequences ....................39 VDE organizer setup menu .................39 6.1.1 Carrying out a test sequence set with VDE Organizer .........40 Custom autotest ....................44...
  • Page 5: General Description

    Metrel d.d. TV 465 General description 1 General description The multifunctional portable appliance tester TV 465 is intended to perform measurements for testing the electrical safety of portable electrical equipment. The following tests can be performed:  earth continuity resistance;...
  • Page 6: Warnings

    Do not use the instrument and accessories if any damage is noticed!  Appliances MUST be fully disconnected from the mains supply before it is connected to the TV 465 in order to perform a PAT test.  Do not touch any test leads/terminals while the appliance is connected to the TV 465.
  • Page 7: Battery And Charging

    If the instrument is not to be used for a long period of time, remove all batteries from the battery compartment.  Alkaline or rechargeable Ni-Cd or Ni-MH batteries (size AA) can be used. Testboy recommends only using rechargeable batteries with a capacity of 2100mAh or higher.
  • Page 8: New Battery Cells Or Cells Unused For A Longer Period

    Cd cells can be subjected to these chemical effects (sometimes called the memory effect). As a result the instrument operation time can be significantly reduced during the initial charging/discharging cycles of the batteries. In this situation, Testboy recommend the following procedure to improve the battery lifetime: Procedure Notes ...
  • Page 9: Standards Applied

    ® Metrel d.d. TV 465 General description 1.4 Standards applied The TV 465 is manufactured and tested in accordance with the following regulations: Electromagnetic compatibility (EMC) EN 61326 Electrical equipment for measurement, control and laboratory use – EMC requirements Class...
  • Page 10: Instrument Description

    ® Metrel d.d. TV 465 Instrument description 2 Instrument description 2.1 Front panel Figure 2.1: Front panel Legend: 128 x 64 dots matrix display with backlight. FAIL Red indicator Indicates PASS/ FAIL of result. PASS Green indicator TEST Starts testing / confirms selected option Selects parameter / changes value of selected parameter.
  • Page 11: Connector Panel

    ® Metrel d.d. TV 465 Instrument description 2.2 Connector panel Figure 2.2: Connector panel Legend: 1 S/EB1 Probe and Earth continuity terminal 2 IEC IEC / Voltage input 3 Protection cover 4 Charger socket For connection of external charger 5 USB connector Communication with PC USB (1.1) port...
  • Page 12: Back Side

    ® Metrel d.d. TV 465 Instrument description 2.3 Back side Figure 2.3: Back side Legend: Inserts for side belt Battery compartment cover Fixing screw for battery compartment cover Back side information label Holder for inclined position of the instrument Figure 2.4: Battery compartment...
  • Page 13: Meaning Of Symbols And Messages On The Instrument Display

    ® Metrel d.d. TV 465 Instrument description 2.4 Meaning of symbols and messages on the instrument display Before performing a measurement, the instrument performs a series of pre-tests to ensure safety and to prevent any damage. These safety pre-tests are checking for any external voltage and load condition on test terminals.
  • Page 14: Battery Indication

    ® Metrel d.d. TV 465 Instrument description WARNING! A high insulation test voltage will be present on the output of the instrument! WARNING! A high insulation test voltage is present on the output of the instrument. Measurement in progress. Test result can be saved.
  • Page 15: Technical Specifications

    ® Metrel d.d. TV 465 Technical specifications 3 Technical specifications 3.1 Earth continuity Range Resolution Accuracy 0.00 Ω ÷ 19.99 Ω 0.01 Ω ±(5 % of reading + 3 digits) Indication only 20.0 Ω ÷ 199.9 Ω 0.1 Ω 200 Ω ÷ 1999 Ω...
  • Page 16: Substitute Leakage Current

    ® Metrel d.d. TV 465 Technical specifications 3.3 Substitute leakage current Range Resolution Accuracy 0.01 mA 0.00 mA ÷ 9.99 mA ±(5 % of reading + 3 digits) 0.1 mA 10.0 mA ÷ 20.0 mA Open circuit voltage ....<50 V AC at rated mains voltage Short circuit current....
  • Page 17: Checkbox Specification

    ® Metrel d.d. TV 465 Technical specifications 3.6 Checkbox specification Instrument Function Reference value Accuracy Earth continuity +/- 1 % of value 0.50 Ω Earth continuity +/- 1 % of value 2.00 Ω Substitute leakage current 5.90 mA +/- 1 % of value Substitute leakage current 0.50 mA...
  • Page 18 ® Metrel d.d. TV 465 Technical specifications Memory..........1500 memory locations The error in operating conditions could be at most the error for reference conditions (specified in the manual for each function) +1 % of measured value + 1 digit, unless otherwise specified in the manual for particular function.
  • Page 19: Main Menu And Test Modes

    ® Metrel d.d. TV 465 Main menu and test modes 4 Main menu and test modes 4.1 Instrument Main menu From the Main menu of the instrument there are five instrument operation modes, Help and Setup that can be selected: Figure 4.1: Instrument Main menu...
  • Page 20: Single Test Menu

    Up to 50 custom autotest sequences can be pre-programmed in this autotest mode. Custom autotest can also be downloaded to and/or uploaded from the PC SW TV 465. Figure 4.4: Custom Autotest menu See chapter 6.2 Custom autotest for detailed description about this test mode.
  • Page 21: Help Menu

    ® Metrel d.d. TV 465 Main menu and test modes 4.7 Help menu Help menu contain schematic diagrams to illustrate how to correctly connect a device under test to the PAT testing instrument. Figure 4.7: Example of help screens Keys: /...
  • Page 22: Language Selection

    PAT tester should be verified at regular intervals and recorded. This is of special importance if the PAT tester is used on a daily basis. The TV 465 includes an in- built Calibration Unit (‘Checkbox’) that is independent from the other instrument’s electronic circuitry and connected directly to the output terminals of the instrument.
  • Page 23 ® Metrel d.d. TV 465 Main menu and test modes Note:  The Checkbox feature should be used to ensure that the meter is reading correctly between calibrations but should not be regarded as a substitute for a full manufacturers calibration on the unit.
  • Page 24 ® Metrel d.d. TV 465 Main menu and test modes Keys: Starts test lead checking procedure. TEST Skips further tests. Checking the IEC test cord The connection for checking the IEC test cord is displayed. Before conducting the check, connect the IEC test cord.
  • Page 25: Test Speed Setup

    ® Metrel d.d. TV 465 Main menu and test modes 4.8.5 Test speed setup In this menu the instrument test speed can be set: Figure 4.17: Test speed menu Options: STANDARD Tests with pauses. No pauses during tests (default). FAST Keys: ...
  • Page 26: User Data

    ® Metrel d.d. TV 465 Main menu and test modes 4.8.7 User data User data can be set in this menu. Figure 4.19: User data menu Keys:  /  Selects the user name. Confirms selection and returns to Setup menu.
  • Page 27: Instrument Data

    ® Metrel d.d. TV 465 Main menu and test modes 4.8.8 Instrument data In this menu the following instrument data is shown:  producer name;  instrument type;  model number;  calibration date;  serial number;  firmware and hardware version.
  • Page 28: Single Test

    ® Metrel d.d. TV 465 Single test 5 Single test In a Single test mode individual tests can be performed. This is especially helpful for troubleshooting. 5.1 Performing measurements in single test mode Select appropriate Single test in Single test Main menu.
  • Page 29: Measurements And Inspections

    ® Metrel d.d. TV 465 Single test 5.2 Measurements and inspections 5.2.1 Visual inspection A thorough visual check must be carried out before each electrical safety test. The following items should be checked:  Inspection of device under test for sign of damage.
  • Page 30: Earth Continuity Resistance

    ® Metrel d.d. TV 465 Single test 5.2.2 Earth continuity resistance This test ensures that the connections between the protective conductor terminal in the mains plug of the device under test and earthed accessible conductive parts of the device under test are satisfactory and of sufficiently low resistance. This test has to be performed on Class I (earthed) appliances.
  • Page 31: Insulation Resistance

    ® Metrel d.d. TV 465 Single test 5.2.3 Insulation resistance The insulation resistance test checks the resistance between live conductors and earthed (or isolated) accessible metal parts of a device under test. This test can disclose faults caused by pollution, moisture, deterioration of the insulation material etc.
  • Page 32: Insulation Resistance - P

    ® Metrel d.d. TV 465 Single test Notes:  When S/EB1 probe is connected during the test then the current through it is also considered.  Consider any warning on the display before starting the measurement!  Do not touch or disconnect the device under test during the measurement or before it is fully discharged! The message »Udisch …«...
  • Page 33: Substitute Leakage

    ® Metrel d.d. TV 465 Single test Figure 5.12: Example of insulation resistance - P measurement results Displayed results: Main result ..... Insulation resistance (LN – P) Notes:  The currents flowing through the PE terminal of the mains test socket will NOT be considered.
  • Page 34: Substitute Leakage - P

    ® Metrel d.d. TV 465 Single test Substitute leakage measurement procedure  Select the SUB. LEAKAGE function.  Set the test parameters.  Connect device under test to the instrument (see figure 5.14).  Press the TEST key for measurement.
  • Page 35 ® Metrel d.d. TV 465 Single test Test circuits for substitute leakage - P measurement Figure 5.17: Measurement of Substitute leakage - P current Substitute leakage - P measurement procedure  Select the SUB. LEAKAGE-P function.  Set the test parameters.
  • Page 36: Polarity Test

    ® Metrel d.d. TV 465 Single test 5.2.7 Polarity test This test checks the polarity of supply cords. The following faults can be detected: L OPEN, N OPEN, PE OPEN, L-N CROSS and MULTIPLE FAULT. Figure 5.19: Polarity test menu Test circuit for polarity test Figure 5.20: Polarity test of IEC cord...
  • Page 37: Functional Test

    ® Metrel d.d. TV 465 Single test 5.2.8 Functional test In its simplest form a functional test is a check to ensure that the appliance is working properly. Note:  This test should only be performed once the appliance has passed all other tests applicable to the device under test.
  • Page 38: Voltage Trms

    Note:  One of the safety features incorporated into the TV 465 is that the voltage measurement will start automatically from any mode when a voltage higher than approximately 50 V (AC or DC) is applied to the IEC connector! If the voltage test has not been preselected then the voltage cannot be stored, but all measurements were prohibited until the voltage is removed.
  • Page 39: Autotest Sequences

    NAME. Note:  The TV 465 instrument does not include all test functions and parameters covered in the Autotest codes. If such a Autotest code is set the instrument automatically perform alternative tests : - a 200 mA Earth continuity test if a test current of more than 200 mA is set.
  • Page 40: Carrying Out A Test Sequence Set With Vde Organizer

    ® Metrel d.d. TV 465 Autotest sequences 6.1.1 Carrying out a test sequence set with VDE Organizer General meaning of keys during a VDE Organizer autotest sequence: /  Sets Organizer option or set value in selected (highlighted) item. Cancels VDE sequence and returns to Main menu.
  • Page 41 ® Metrel d.d. TV 465 Autotest sequences Insulation resistance measurement This test is offered if it is applicable according to the VDE Organizer setting. The measurement is described in chapter 5.2.3 Insulation resistance. Figure 6.5: VDE Organizer – Insulation resistance starting screen...
  • Page 42 ® Metrel d.d. TV 465 Autotest sequences Insulation resistance - P measurement This test is offered if it is applicable according to the VDE Organizer setting. The measurement is described in chapter 5.2.4 Insulation resistance - P. Figure 6.9: VDE Organizer autotest – Insulation resistance - P starting screen...
  • Page 43 ® Metrel d.d. TV 465 Autotest sequences Functional test Figure 6.13: VDE Organizer – Functional result screens Options in Functional test result screen: To be applied manually. PASS / FAIL Option if IEC leads, multiple outlets without electronic parts is set: Starts Polarity test.
  • Page 44: Custom Autotest

    6.2 Custom autotest In the autotest custom menu user-defined autotest procedures can be performed via the PC SW TV 465. Up to 50 custom autotest sequences can be pre-programmed in this autotest mode. Commonly used pre-programmed autotest sequences are added to the list by default.
  • Page 45: Code Autotest

    ® Metrel d.d. TV 465 Autotest sequences 6.4 Code Autotest Code Autotest menu supports operation with pre-defined test codes, barcodes and RFID tags. The instrument supports the following functions:  manual selection of pre-defined autotest shortcut codes;  reading pre-defined autotest shortcut codes from barcode labels;...
  • Page 46: Carrying Out (Code, Simple And Custom) Autotest Sequences

    ® Metrel d.d. TV 465 Autotest sequences Reading appliance ID number with barcode reader or RFID reader/writer When the instrument is in the Save results menu, appliance ID can be scanned from a barcode label with the barcode reader or can be read from RFID tag with the RFID reader/writer.
  • Page 47: Insulation Resistance Measurement

    ® Metrel d.d. TV 465 Autotest sequences Figure 6.20: Earth continuity result screen Options in Earth continuity result screen: Proceeds to next step. NEXT Repeat the test (use in case of multiple earthed points). Highest result will REPEAT be stored.
  • Page 48: Insulation Resistance - P Measurement

    ® Metrel d.d. TV 465 Autotest sequences 6.5.5 Insulation resistance - P measurement The test is offered if it is applicable according to the autotest setting. The Insulation resistance - P starting screen is displayed first. Measurement and options in Insulation resistance - P starting screen are described in chapter 5.2.4 Insulation resistance - P.
  • Page 49: Polarity Test

    ® Metrel d.d. TV 465 Autotest sequences 6.5.7 Polarity test The test is offered if it is applicable according to the autotest setting. The Polarity test starting screen is displayed first. Measurement and options in Polarity test starting screen are described in chapter 5.2.7 Polarity test.
  • Page 50: Handling Autotest Results

    ® Metrel d.d. TV 465 Autotest sequences 6.6 Handling autotest results After the Code, Simple or Custom autotest is finished, the Main autotest result screen will be displayed including an overall  /  indication. Figure 6.32: Main autotest result screen Options in Autotest results screen: Views individual results.
  • Page 51: Working With Autotest Results

    ® Metrel d.d. TV 465 Working with autotest results 7 Working with autotest results 7.1 Saving autotest results After selecting Save results in Autotest results menu, the autotest results will be stored in the internal memory of the instrument. The appliance ID number and NAME can be added to the test results before the results are saved: Figure 7.1: Save results menu (Appliance ID)
  • Page 52: Recalling Results

    ® Metrel d.d. TV 465 Working with autotest results 7.2 Recalling results Saved autotest results can be recalled, printed or deleted from the Memory menu. Enter the Memory menu from the Setup menu. Figure 7.4: Memory menu To enter the Recall results menu select Recall results in Memory menu. A list of Appliance ID’s and NAMES are displayed in a chronological order (last performed measurement is...
  • Page 53: Deleting Individual Autotest Results

    ® Metrel d.d. TV 465 Working with autotest results 7.3 Deleting individual autotest results To enter the Delete results menu select Delete results in Memory menu. A list of Appliance ID’s and NAMES are displayed in a chronological order (last performed measurement will be displayed at the top of the list).
  • Page 54: Printing And Rfid Tagging Of Individual Autotest Results

    ® Metrel d.d. TV 465 Communication 7.5 Printing and RFID tagging of individual autotest results To print labels or results and write RFID tags select Print data / RFID in Memory menu. A list of Appliance ID’s and NAMES are displayed in a chronological order (last performed measurement will be displayed at the top of the list).
  • Page 55: Communication

    ® Metrel d.d. TV 465 Communication 8 Communication The instrument can communicate with the PC SW TV 465 PC software. The following actions are supported:  Saved results can be downloaded and stored to a PC.  Checkbox results can be downloaded and stored to the PC.
  • Page 56: Maintenance

    The calibration should be done by an authorized technical person only. 9.2 Service For repairs under or out of warranty please contact your distributor for further information. Unauthorized person is not allowed to open the TV 465 instrument. There are no user replaceable parts inside the instrument. 9.3 Cleaning Use a soft cloth, slightly moistened with soapy water or alcohol to clean the surface of the instrument.
  • Page 57: Instrument Set And Accessories

    ® Metrel d.d. TV 465 Instrument set and accessories 10 Instrument set and accessories Standard set of the instrument  Instrument TV 465  Small soft carrying bag  Test probe (black)  Crocodile clip (black)  Test lead (1.5 m, black) ...
  • Page 58: Appendix A - Barcode Formats

    Appendix A – Barcode formats Appendix A – Barcode formats The instrument TV 465 supports two barcode formats (single and double). Autotest code and appliance ID Autotest codes are represented as a three digit code. These autotest codes can also be represented by the barcode.
  • Page 59: Appendix B - Pre-Programmed Autotests (Ger)

    ® Metrel d.d. TV 465 Appendix B – Pre-programmed autotests (GER) Appendix B – Pre-programmed autotests (GER) Pre-programmed autotest sequences table Autotest shortcut code Kl_1_Iso Kl_1_Iso_BLT Kl_1_Ia Kl_1_Ia_BLT     Visual test Output 200 mA 200 mA 200 mA...
  • Page 60 ® Metrel d.d. TV 465 Appendix B – Pre-programmed autotests (GER) Pre-programmed autotest sequences table (cont’d) Autotest shortcut code Kl_2_Iso Kl_2_Ibs Kl_1_IsoIa Kl_1_IsoIaBLT     Visual test   Output 200 mA 200 mA Earth   Limit 0.30 Ω...
  • Page 61 ® Metrel d.d. TV 465 Appendix B – Pre-programmed autotests (GER) Pre-programmed autotest sequences table (cont’d) Autotest shortcut code Kl_2_IsoIbs Kl_2 Kl_3_Iso Kl_3     Visual test     Output Earth     Limit continuity ...
  • Page 62 ® Metrel d.d. TV 465 Appendix B – Pre-programmed autotests (GER) TESTBOY GmbH VDE tester test type card Code Autotest sequence name and descriptions Limits Barcode Testing according to VDE. Class 1 device. Earth bond: 0.30 Ω Insulation resistance substitute Kl_1_Iso Insulation: 1.00 MΩ...
  • Page 63 ® Metrel d.d. TV 465 Appendix B – Pre-programmed autotests (GER) TESTBOY GmbH VDE tester test type card (cont'd) Testing according to VDE. Kl_2 Class 2 device without any isolated accessible conductive parts. A1 0 Testing according to VDE. Insulation - P: 0.25 MΩ...
  • Page 64: Appendix C - Pre-Programmed Autotests (Nl)

    ® Metrel d.d. TV 465 Appendix C – Pre-programmed autotests (NL) Appendix C – Pre-programmed autotests (NL) Pre-programmed autotest sequences table Autotest shortcut code Kl_1_ALG Kl_2_ALG Kl_1_HEATERS KL_1_PC     Visual test  Output 200 mA 200 mA...
  • Page 65 ® Metrel d.d. TV 465 Appendix C – Pre-programmed autotests (NL) Pre-programmed autotest sequences table (cont’d) Autotest shortcut code Kl_3_ALG Kl_1_AGMD KABEL_5M_2.5MM KABEL_15M_2.5MM     Visual test  Output 200 mA 200 mA 200 mA Earth  Limit 0.30 Ω...
  • Page 66 ® Metrel d.d. TV 465 Appendix C – Pre-programmed autotests (NL) Pre-programmed autotest sequences table (cont’d) Autotest shortcut code KABEL_25M_2.5M KABEL_50M_2.5M   Visual test Output 200 mA 200 mA Earth Limit 0.70 Ω 1.00 Ω continuity Time Output 500 V...
  • Page 67: Appendix D - Simple Test Codes

    ® Metrel d.d. TV 465 Appendix D – Simple test codes Appendix D – Simple test codes Simple test codes CLASS. I CLASS. II CLASS. III    Visual test   Output 200 mA   Earth continuity Limit 0.30 Ω...
  • Page 68 ® Metrel d.d. TV 465 Appendix D – Simple test codes...

Table of Contents