Wavetek 144 Instruction Manual page 30

Hf sweep generator
Table of Contents

Advertisement

9.
Set the
voltage
between
IC3(2),
and
IC3(3) to
0
Vdc
±100
/iV,
using
R
107.
See
Figure
4-4.
10.
Set the voltage
between
IC4(2),
and
IC4(3)
to
0
Vdc
±100
juV,
using
R
1
13.
11.
Set the voltage
between
IC5(2),
and
IC5(3) to
0
Vdc
±1
00
juV,
using
R123.
12.
Set the voltage
between
IC6(2),
and
IC6(3) to 0
Vdc
±100
/LiV,
using
R125.
13.
With
FREQ
DIAL
still
fully
clockwise,
set
the voltage
between
IC2(2)
and ground
(at
R95)
to
0
Vdc
±100
jLtV,
using
R93.
14.
Set
FREQ
DIAL
to
1
and
adjust voltage
between
TP6
and
TP7
to
0
Vdc ±200
/iV,
using
R129.
15.
Adjust
voltage
between
TPS
and
TP9
to
0
Vdc
±200
jLxV,
using
R132.
4.2.5 Triangle
Amplitude Adjustment
16.
Set
FREQ
HZ
to
X
100,
FREQ
DIAL
to 10,
WAVE-
FQRM
SELECTQR
to
TRIANGLE,
and
SYMMETRY
to
OFF.
17.
Connect
the
oscilloscope
probe
(a
type
W
or
1A5
plug-in
module
should be used with the scope)
to
\
position of the
WAVEFORM
selector switch,
SW2,
see
Figure
4-4.
18.
Adjust
the
triangle
amplitude
to
±1.25
V
±5
mV,
using
R12
and R23,
see
Drawing
142-010.
4.2.6
Sine Distortion
19.
Set
FREQ
HZ
to
X
100,
VERNIER
to
CAL,
WAVE-
FORM
SELECTOR
to
SINE,
FREQ
DIAL
to
10,
and
OUTPUT ATTEN
to
0
dB.
20.
Set
30
V
P-P
MAX
control to
3/4ths
of
full
CW,
and connect
oscilloscope
probe
to
^
position of
the
WAVEFORM
selector
switch (SW2),
see
Figure
4-4.
21.
Adjust
R153
so that
the
dc
offset
of the
sine
ampli-
fier
is
zero.
22.
Connect
the
test
equipment
as
shown
in
Figure
4-1,
and
adjust
R159, R168, and R 145
for
minimum
sine
wave
distortion.
Figure
4-1
-
Distortion Analysis
Test
Set-Up
23.
Set
FREQ
HZ
to
X
IK
and
readjust
R159, R168,
and
R
1
45
for
optimum
sine distortion.
24.
Set
FREQ HZ
to
X
100 and
readjust
R159, R168,
and
R145
if
necessary
25.
Alternately
perform
steps
24
and 25
until
the
least
amount
of
sine
wave
distortion
is
obtained
on
both
ranges
typically
.14%.
26.
Adjust
R165
for a 2.5
V
p-p
signal
output from
the
sine
amplifier
±25
mV.
27.
Readjust
R153
so
that the
dc
offset
of the
sine
amplifier
is
0
V
±5
mV.
28.
Connect
the
Spectrum
Analyzer
as
shown
in
Figure
4-2,
and check
the
sine
wave
distortion
at
frequencies
higher
than
600
kHz.
If
distortion
does not
meet
the
specification,
check
calibration
of
C95,
follow proce-
dure
4.2.10.
Figure 4-2
-
Spectrum
Analysis Test
Set-Up
4-3
7/71

Advertisement

Table of Contents
loading

Table of Contents