Wavetek 144 Instruction Manual page 18

Hf sweep generator
Table of Contents

Advertisement

Figure 3-4
Sweep
Generator, Triggered or
Gated
Modes
When
this
instrument
is
in
the
TRIG mode
and
ready,
the
Main Generator Diode Gate
is
switched
off
by
the
Trigger/Gate Control,
and
the
Timing
Capacitor
is
dis-
charged through the shunt
path.
When
a
trigger signal
arrives at
the Trigger/Gate Control, the control
flip-flop
changes
the
state,
opening the Diode
Gate.
Current flows
into the
Timing
Capacitor, allowing the passive
integrator
to generate a
triangle
waveform.
The
control
flip-flop
is
reset
by
the square
wave from
the
Sync
Amplifier
when
the
proper phase
triggers
the
flip-flop.
The
flip-flop
output,
in
conjunction with the
level-shift circuit,
closes
the
Diode
Gate.
Thus, the
single
cycle generation of
a selected
wave-
form
is
accomplished.
When
in
the
GATED
mode,
the
output
of the Trigger
Level
Comparator/Squaring
circuit
is
coupled
directly to
the Trigger/Gate Control.
The
sequence of operation
is
the
same
as that
of the
TRIG
mode,
except
that
one
or
more
cycles are
generated for the duration of the
trigger
(gate) signal
applied to the
TRIG
IN connector.
SWEEP
MODES
In
the
SWEEP
modes
of operation,
a
third
input
(the
Sweep
Generator output)
is
added
to the
VCG
Amplifiers through
the
SWEEP WIDTH
control.
In
the
CONT
SWP
mode,
the
Sweep
Generator and the Main Generator
run continuously.
In
the
TRIG
SWP
and
SWP
HOLD
modes,
the
Main
Genera-
tor
runs continuously
at
the
frequency
set
by
the frequency
controls
until a trigger
or gate pulse
appears
at
the
TRIG
IN
connector.
(a)
Continuous
Sweep
Mode
Refer to Figure
3-5. In
the
CONT SWP
mode,
the
basic
frequency of the
Main
Generator
is
set
by
the
FREQ
3-4
8/73