Running The Diagnostics - Labconco SteamScrubber 45780 Series User Manual

Glassware washers
Table of Contents

Advertisement

If the washer is interrupted in the middle of the diagnostic steps, wait 4 to 5
seconds before opening the washer door to prevent hot water from splashing out.
The hot water and steam from the washer may cause scalding.

Running the Diagnostics

To ensure your washer is operating properly, perform the diagnostics procedure
detailed below, which will take approximately 10 minutes to complete. Refer to
the figure below to familiarize yourself with the control panel before you begin
the procedure.
To run the diagnostics test:
1. Press the ▲ button on the control panel and close the door and slide the door
latch to the right. Hold for 5 seconds.
2. The display will show the Diagnostics screen and the indicator on the left side
will point to TAP WATER VALVE.
3. Press either the ► or ◄ button to turn the tap water valve ON. Tap water
should enter the washer. Allow water to enter for approximately 1 minute.
Press either the ► or ◄ button to turn the water off.
4. Press the ▼ button and the arrow on the display will point to PURE WATER.
5. Press either the ► or ◄ button turn the pure water pump ON and open the
pure water valve. Allow pure water to enter the washer. After about 3
minutes the fill switch should cause the water flow to turn off.
6. Press the ▼ button and the arrow on the display will point to WASH PUMP.
7. Press the ► or ◄ button and the wash pump should energize.
Product Service 1-800-522-7658
Chapter 3: Getting Started
21

Hide quick links: