5. Reprogram the pulse on time and/or the pulse off time as follows:
Pulse on time
a. Turn power off
b. Turn power on while holding pulse "On" button. The LED display for Speed
displays current pulse-on time. The pulse "On" button can be released if
necessary once the display lights.
c. Change the pulse-on time using the up/down arrows to the right of the
illuminated Speed display.
d. When the power is turned off, the value on the display is stored as the
new pulse "On" time.
IMPORTANT: This unit is designed for intermittent use only. Do not operate this unit for more than two hours at any one time.
A
CCESSORIES
P
N
.
ART
O
58816-307
58816-308
58816-330
58816-363
58816-396
58816-306
(C
)
ONTINUED
D
ESCRIPTION
96 well microassay (gray)
10mm tube rack (gray)
12mm tube rack (blue)
13mm tube rack (yellow)
16mm tube rack (green)
Replacement tray pad set (gray)
Pulse off time
a. Turn power off
b. Turn power on while holding pulse "Off" button. The LED display for Timer
displays current pulse off time. The pulse "Off" button can be released if
necessary once the display lights.
c. Change the pulse-off time using the up/down arrows to the right of the
illuminated Timer display.
d. When the power is turned off, the value on the display is stored as the
new pulse "Off" time.