Saving Event Logs - ION ARA H2S-H Instrument User Manual

Single gas detector
Table of Contents

Advertisement

In order to communicate, the ARA needs to be in front of the IR Link.
When the ARA is in position, click 'Read Device'.
The settings will be retrieved from the device and displayed within the software.
NOTE:
If the communication fails, the following message will be displayed "ARA Connect ERROR".
Simply adjust and reposition the ARA in relation to the IR Link. If you have tried this a few times and haven't
had any success, please contact ION Science.
Once the instrument is connected and has successfully been read, the settings can be changed. Click on Write
Device to save the settings to the ARA.

Saving Event Logs

The ARA stores the last 30 events. To retrieve and save the events use IR Command and connect the ARA using
IR Link then select 'Save Event Log'.
This will then transfer the last 30 events to a spreadsheet and will contain the following information.
Serial Number
Firmware Version
User ID
Life Remaining
Number of events
Duration of events
Number of self-tests
Date
Time
Bump test
Duration
Reading
Alarm
Pioneering Gas Sensing Technology.
ARA Instrument User Manual V1.1
Page 20 of 27
ionscience.com