Maintenance And External Care Precautions; Transport And Storage Precautions; Installation; Splice Overview - FIBER FOX MINI3S User Manual

Arc fusion splicer
Table of Contents

Advertisement

Fumes, bad smell, abnormal noise or over heat.
Liquid or other matter falls into cabinet
The splicer is damaged or dropped.
If any of these faults occurs, please contact our service center immediately. Leaving
the splicer in a damaged state without any prompt measures may cause equipment
failure, electric shock, or fire and may result in injury or death.
⑧ Do not use compressed gas or canned air to clean the splicer. They may contain
flammable materials that could ignite during the electrical discharge.
⑨ Please use MINI3S specific standard AC adapter only. Using an improper AC
power source may cause fuming, electric shock or equipment damage and may even
result in fire, injury or death.
⑩ Please use MINI3S specific AC power cord only. Do not place any heavy objects
on the AC power cord. Keep the power cord away from heat source. Do not modify
the power cord. Using an improper cord or a damaged cord may cause fuming,
electric shock or equipment damage and may even result in fire, injury or death.

2.3 Maintenance and External Care Precautions

① Always avoid using hard objects to clean V-grooves and electrodes.
② Always avoid using acetone, thinner, benzol or alcohol when cleaning any part of
the splicer, except for the places advised.
③ Use a dry cloth to remove dust and dirt from the splicer.
④ If the outside of the splicer is dirty, plunge a soft cloth into diluted neutral
washing up liquid, wring out the cloth and clean. Dry the splicer with a dry cloth but
DO NOT use furniture polish or other cleaning agents.
⑤ Always follow the maintenance instructions in this manual.

2.4 Transport and Storage Precautions

① When the splicer is moved from cold to warm environment, you should allow the
splicer to warm up gradually. Otherwise, the condensation generated inside will
bring harmful effects to the splicer.
② Pack the fusion splicer well for long time storage.
③ Keep the splicer clean and dry.
④ The splicer is precision adjusted and aligned. Always keep the splicer in its
carrying case to protect from damage and dirt. Put cushion package outside the
carrying case for long distance transportation.
⑤ Always avoid leaving the splicer in direct sunlight or expose to excessive heat.
⑥ Keep the humidity to a minimum level where the splicer is stored. The humidity
must not exceed 95%.
08

2.5 Installation

Important! Follow these instructions carefully.
Unpacking the Splicer
Hold the handle upwards, and then lift the splicer out of the carrying case. As shown
below.

2.6 Splice Overview

Heat
er
Handle
Battery
Indicator
off button
On/
Display
09

Advertisement

Table of Contents
loading

Table of Contents