Low Frequency Multiplexer Modules

SCPI Programming Examples for the Multiplexer Modules

The programming examples below provide you with SCPI command examples to
use for actions specific to the MUX modules.
The slot and channel addressing scheme used in these examples follow the form
sccc where s is the mainframe slot number (1 through 8) and ccc is the
three-digit channel number. For information on specific MUX channel
configurations, refer to the simplified schematics contained in each MUX section
of this manual.
For complete information on the SCPI commands used to program the 34980A,
with example programs, refer to the Keysight 34980A Programmer's Reference
contained on the 34980A Product Reference CD.
Opening and Closing Channels
Example - Closing and opening channels on the armature and reed MUX
modules:
any channel in a bank is defined to be part of the scan list, and a scan is
occurring, attempting to close another channel (including Analog Bus channels)
within the same bank will result in an error. Channel closures in the other bank
are allowed as long as no channels are part of the scan list.
The following commands close and open channels 13 and 15 through 18 in slot
3.
ROUTe:CLOSe (@3013,3015:3018)
ROUTe:OPEN (@3013,3015:3018)
Example -Closing channels on the FET MUX module: The FET MUX module
supports a 1:N type closure, meaning that you can have only one channel per
bank closed at a time. The following command closes then automatically opens
each channel from 1-19 (Bank 1) in succession, leaving channel 20 closed. Then
the command continues closing and opening channels 21 to 39 (Bank 2), then
leaving channel 40 closed.
At the end, only channels 20 and 40 will be closed, while all other channels will
have been closed and then opened. In this process,
a channel will open before the next channel in succession closes, making this a
"break-before-make" series.
ROUTe:CLOSe (@3001:3040)
The following command opens the closed channel on Bank 1 of a FET MUX
module in slot 3, and closes channel 15 on that bank.
ROUTe:CLOSe (@3015)
12
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SCPI Programming Examples for the Multiplexer Modules
This command closes the specified channels on a MUX module. If
Keysight 34921A-34925A Low Frequency Multiplexer Modules User's Guide