Positioning the ADTS

CAUTION
TO OPERATE, PLACE THE ADTS ON A HORIZONTAL SURFACE WITH THE FRONT PANEL UPPER-
MOST, THIS ALLOWS THE WATER IN THE WATER FILTER TO VENT. WATER CAN CONTAMINATE
THE ADTS MANIFOLD AND AFFECT ADTS PERFORMANCE.
: In control mode, the water drain, located on the left-hand side of the unit near the cooling vent,
NOTE
produces a flow of air and some water. The amount of water depends on the humidity and the oper-
ating time in control mode.
It is important that the position of the ADTS in relation to the aircraft altitude sensors is known. An
altitude correction must be made to allow for the difference in height between the ADTS reference
level and the reference level of the aircraft's altitude sensors (refer to the Aircraft Maintenance
Manual for this information).
Enter the altitude correction value, see section Settings.
CAUTION
OBSERVE THE APPROPRIATE SAFETY INSTRUCTIONS AND TESTING PROCEDURES DETAILED
IN THE AIRCRAFT MAINTENANCE MANUALS AND COMPONENT MAINTENANCE MANUALS.