Positioning of the ADTS 2XX

WARNING:
DO NOT DISCONNECT THE ADVANCED HAND TERMINAL WHEN ENERGIZED IN THE
HAZARDOUS AREA. THIS CAN CAUSE AN EXPLOSION.
To operate safely, the ADTS 2XX must be placed outside the user defined zone 2 hazardous area.
Only the ATEX certified hand terminal may be used inside the defined zone 2 hazardous area (refer
to section 5 for a definition).
C
:
AUTION
T
,
O OPERATE
PLACE THE UNIT ON A HORIZONTAL SURFACE WITH
THE FRONT PANEL UPPERMOST
1
CONNECTOR PS
WATER FILTER TO VENT
CONTROLLER MANIFOLD AND AFFECT CONTROLLER PERFORMANCE
Note:
In control mode, the water drain, located near the cooling vent, produces a flow of air and some water.
The amount of water depends on the humidity and the operating time in control mode.
• It is important that the position of the ADTS 2XX in relation to the aircraft altitude sensors is known. An
altitude correction must be made to allow for the difference in height between the ADTS reference level
and the reference level of the aircraft's altitude sensors (refer to the Aircraft Maintenance Manual for this
information).
• Enter the altitude correction value in the SET-UP menu detailed in section 3.
WARNING:
OBSERVE THE APPROPRIATE SAFETY INSTRUCTIONS AND TESTING PROCEDURES DETAILED
IN THE AIRCRAFT MAINTENANCE MANUALS AND COMPONENT MAINTENANCE MANUALS.
(A)
OR WITH THE PNEUMATIC
(B)
UPPERMOST
THIS ALLOWS THE WATER IN THE
. W
ATER COULD CONTAMINATE THE
reference level
positive
altitude
correction
value
reference level
F
2-2 ADTS 2XX A
IGURE
Druck ADTS 2XX User Manual
A
.
inst
AIR DATA
COMPUTER
C
LTITUDE
ORRECTION
B
ref
vent
K0451 Issue No. 1