2.
Press
to select Testing Menu.
ENTER
3.
Scroll to and select 008 Ammonia-N HR
from menu.
4.
Rinse a clean tube (0290) three times
with the ammonia nitrogen standard.
Fill to the 10 mL line with the ammonia
nitrogen standard. Cap the tube. Dry the
tube with a lint-free cloth.
5.
Insert tube into chamber, close lid and
select Scan Blank.
6.
Remove tube from colorimeter. Add 8
drops of Ammonia Nitrogen Reagent #1
(V-4797). Cap and mix.
008 Ammonia-N HR
12:00:00
008 Ammonia-N HR
Scan Blank
Scan Sample
12:00:00
7
All Tests
001/500
001/500