DC1500 AMMONIA NITROGEN KIT

„ KIT CONTENTS

QUANTITY
CONTENTS
30 mL
Ammonia Nitrogen Reagent #1
3 x 30 mL
*Ammonia Nitrogen Reagent #2
1
Pipet, 1 mL Plastic
1
Colorimeter Tubes, w/caps,set of 6
1
Water Sample Collecting Bottle
1
1500 Colorimeter for Ammonia Nitrogen
1
USB Wall Adapter
1
USB Cable
1
DC1500-AM Ammonia Colorimeter, Manual
1
DC1500-AM Ammonia Colorimeter, Quick Start Guide
*Reagent is a potential health hazard. READ SDS:
lamotte.com. Emergency information:
Chem-Tel USA 1-800-255-3924
Int'l, call collect, 813-248-0585
To order individual reagents or test kit components, use the specified code number.

„ ACCESSORIES

DESCRIPTION
Test Tubes, with Caps
Replacement Chamber
USB Cable
USB Wall Adapter
Car Charger
TEST METHODS SPECIFICATIONS

„ INTRODUCTION

Ammonia nitrogen is present in various concentrations in many surface and ground water
supplies. Any sudden change in the concentration of ammonia nitrogen in a water supply
is cause for suspicion. A product of microbiological activity, ammonia nitrogen is sometimes
accepted as chemical evidence of pollution when encountered in natural waters.
Ammonia is rapidly oxidized in natural water systems by special bacterial groups that
produce nitrite and nitrate. This oxidation requires that dissolved oxygen be available in
the water. Ammonia is an additional source of nitrogen as a nutrient which may contribute to
the expanded growth of undesirable algae and other forms of plant growth that overload the
natural system and cause pollution.

„ APPLICATION

Drinking, surface, and saline waters; domestic and industrial wastes.

„ RANGE

0 to 5.0 ppm Ammonia Nitrogen

„ METHOD

Ammonia forms a colored complex with Nessler's Reagent in proportion to the amount of
ammonia present in the sample. Rochelle salt is added to prevent precipitation of calcium or
magnesium in undistilled samples.
„ HANDLING & PRESERVATION
Preservation is accomplished by the addition of 2 mL of concentrated H

„ INTERFERENCES

Sample turbidity and color may interfere. Turbidity may be removed by a filtration procedure.
Color interference may be eliminated by adjusting the instrument to 100%T with a sample blank.
CODE
WARNING: Only use the USB cable (1720-01) that is
supplied with the kit. Make no substitutions.
0290-6
3-0038
1720-01
1721
5-0132
3
CODE
V-4797-G
*V-4798-G
0354
0290-6
0688
27926-AM
1721
1720-01
3241-MN
3241-QG
SO
at 4°C.
2
4

Advertisement

Table of Contents
loading

This manual is also suitable for:

3241

Table of Contents