10-3 Details of the Function Table

Class
Item and Parameter
Cond
Condition
5t-b
Stability band width
Hold
Hold function
trc
Zero tracking
5pd
Display refresh rate
pnt
ba5fnc
Decimal point
Environment
Display
p-on
Auto display-ON
poff
Auto display-OFF
g5i
Capacity indicator
f1l
Filter
p-tr
Zero upon power-on
5pn
Span calibration
Cl adj
Clock
Cp
Comparator mode
Cp fnc
Comparator
Cp in
Input method
Fast response rate, sensitive value FAST
0
1
Slow response rate, stable value SLOW
2
Stable range is ±1 digit
0
1
Stable range is ±3 digits
2
0
OFF
1
ON
0
OFF
1
Normal
2
Strong
3
Very strong
0
5 times/second
1
10 times/second
0
Point (.)
1
Comma (,)
0
OFF
1
ON
0
OFF
1
ON (10 minutes)
0
OFF
1
ON
0
Not used
1
Used (when the balance is used as a mass comparator)
0
Sets the display to zero.
Does not set the display to zero. Displays the previous value.
1
With this setting, do not perform re-zero operations frequently.
Refer to "Zero upon power-on" on page 37.
0
Performs zero and span calibration
Performs span calibration only.
Span calibration using the internal mass is possible with a tare
1
on the weighing pan.
Refer to "Span calibration" on page 37.
Refer to the GX series instruction
manual, "10-9 Clock and Calendar
Function".
0
No comparison
1
Comparison, excluding "near zero" when stable or overloaded
2
Comparison, including "near zero" when stable or overloaded
3
Continuous comparison, excluding "near zero"
4
Continuous comparison, including "near zero"
0
Digital input, upper/lower limits
Weighing input, upper/lower
1
limits
2
Digital input, reference value
3
Weighing input, reference value
Factory setting
Note: "Digit" is a unit of minimum weighing value
33
Description
Can be changed by response
MID.
adjustment. With "hold 1",
sets the averaging time.
The stabilization indicator illuminates
when the display fluctuation is
within the range. With "hold 1",
sets the averaging time.
Holds the display when stable
in animal mode. With "Hold 1",
ANIMAL turns on.
Keeps zero display by tracking
zero drift.
Period to refresh the display
Decimal point format
Turns on the weighing mode
display when the AC adapter
is plugged in.
Turns off the display after 10
minutes of inactivity.
Capacity indicator
Zero: 0%,
Maximum capacity: 100%
Confirms and sets the time and
date. The time and date are
added to the output data.
Cp Hi, Cp lo can be selected.
Cp ref, Cp lmt can be
selected..