Calibrate Screen (F
For ME14 and PE14 Meters
Recalibrates the meter for different fluids.
Table 1. Calibration Factors
Fluids
Oil (10W - 30)
Gear Lube
Automatic
Transmission Fluid
Antifreeze
•
Calibration factors at 70°F (21°C) at 2.0 gpm (7.6 lpm).
•
Calibration number may vary slightly due to temperature
or flow rate.
Calibrate Screen 1
L
Calibrate Screen 2
M
L
F
. 14
IG
312865A
. 14)
IG
Calibration Number
Quarts
Liters
177
187
177
187
177
187
157
165
J
D
ti12173a
K
ti12174a
ME14 and PE14 include D - M
D. Wrench Icon: Returns user to Main Screen.
J. Calibration Factor: Displays amount operator
dispensed during meter calibration.
K. Preset Calibration Factor: Preset calibration
amount displays.
Adjusting the Preset Calibration Factor:
1. Use your LEFT or RIGHT ARROW to move the
curser to the Calibration Factor Field (K) .
2. Use the UP or DOWN ARROW to manually adjust
the Preset Calibration Factor up or down (this can
only be done on Calibrate Screen 1 (F
3. Press center, ENTER button to confirm new setting.
New setting displays in Calibration Factor field (J).
L. Start/End button: Clicked at the start and completion
of calibration. See Recalibrating the Meter (page 14).
M. Dispense: Only appears on Screen 2 (F
Displays amount operator dispensed during calibration
test.