Specifi Cation - Konica Minolta PULSOX-300 Instruction Manual

Oxygen saturation monitor
Hide thumbs Also See for PULSOX-300:
Table of Contents

Advertisement

Specifi cation
■ Dual-wavelength pulse type (665nm/880nm) arterial blood oxygen satura-
tion measuring instrument
Name:
Model name: PULSOX-300
■ Functions:
Measuring range:
Accuracy:
✻ PULSOX-300 has been clinically validated for use on adults.
■ Display
Display type: Liquid crystal display
Oxygen saturation (SpO
Pulse rate number
Pulse level meter (5 steps)
Battery indication
Error messages indication
■ Warning functions
SpO
: Low SpO
2
Battery indication (fl ashing when the battery is nealy empty)
Error message indication
■ Operating conditions
temperature/humidity range
0 to 40°C (32 to 104°F); 30 to 85% relative humidity with no condensation
atmospheric pressure/altitude range
700 to 1060hPa (altitude: –400 to 3000m)
■ Storage conditions
temperature/humidity range
–10 to 60°C (14 to 140°F); 10 to 95% relative humidity with no condensation
atmospheric pressure/altitude range
700 to 1060hPa (altitude: –400 to 3000m)
■ Power
1 AAA-size battery: 1.5V
approx. 16 hours with alkaline battery)
■ Dimensions (W × H × D)
68 × 58 × 15 mm
■ Weight
56g (including battery/excluding wrist band and probes)
■ Usable life
6 years [Verifi ed by KONICA MINOLTA SENSING, INC. (based on own data)]
Get other manuals https://www.bkmanuals.com
Oxygen Saturation Monitor
SpO
:
2
Pulse rate: 30 to 230 bpm
SpO
:
2
Pulse rate: ±2 bpm (30 to 100 range)
)
2
warning (fl ashing display)
2
100mW (Service life under continual usage:
0 to 100% SpO
2
±2% SpO
(70 to 100% range, 1S.D.)
2
or ±2% of value (100 to 230 range)
23

Advertisement

Table of Contents
loading