Konica Minolta PULSOX-300 Instruction Manual page 11

Oxygen saturation monitor
Hide thumbs Also See for PULSOX-300:
Table of Contents

Advertisement

When influenced by abnormal hemoglobin like carbon
monoxide hemoglobin (HbCO) and methemoglobin
When pigments such as cardiogreen, intravascular dyes
and indocyanine greenare present in the blood
When fi nger nails are polished
When fi nger nails are polished
Tables 1 and 2 below show the errors which may occur due to
these hemoglobins.
Difference from the SaO
<Table 1>
SaO
2
50
-0.1
70
-0.1
90
-0.2
100
-0.2
Measured value is always
lower than the actual value.
Get other manuals https://www.bkmanuals.com
HbCO
1%
5%
10%
-0.7
-1.5
-0.7
-1.5
-0.8
-1.6
-0.8
-1.7
value to displayed value
2
<Table 2>
Methemoglobin
SaO
2
1%
50
+0.2
70
-0.6
90
-1.5
100
-1.8
Displayed value is sometimes
higher than the actual value if
SaO
is around 50%.
2
5%
10%
+1.3
+3.2
-2.3
-3.2
-6.0
-9.6
-7.5
-12.2
9