Battery Removal And Replacement; Battery Life; Figure 5: Battery Removal And Replacement - Welch Allyn SureTemp 986 Service Manual

Table of Contents

Advertisement

2.3

Battery Removal and Replacement

2.3.1 Battery Life

Model 986 Thermometer

Figure 5: Battery Removal and Replacement

1. Remove the Probe Cover box holder on the back of the
instrument by pressing with your index finger on the latch
located on the rear of the Probe Cover box holder.
2. Remove the batteries.
3. Press the Mode button for approximately five seconds to
discharge the electronics.
4. Install two new AAA, alkaline batteries according to the
battery polarities marked inside the battery compartment.
Verify that the thermometer completes a power on reset
test (page 7), and then goes blank.
CAUTION: Incorrect battery polarity may result in damage
to the thermometer.
5. Snap the probe cover storage assembly in place by inserting the
bottom tabs first and then snapping the latch on top.
Under normal use, battery life will provide approximately 3,000
normal temperature measurements, based on an average 22.2°C
(72.0°F) ambient temperature. Colder ambient temperatures,
excessive peak monitor measurements, and other usage patterns
can reduce battery life.
Instruments are shipped with fresh batteries, but we cannot
guarantee full life from the first set of batteries due to potential long
storage times between shipping and actual use. Battery life can
also be reduced by storage at elevated temperatures.
If you are experiencing short battery life, refer to Battery Life
Problems in the Troubleshooting table on page 22.