[SENSe:]TCAPture[1|2]:RANGe
Specifies range value for time capture data.
Command Syntax:
[SENSe:]TCAPture[1|2]:RANGe <number>|<bound>
<number>
<bound>
Example Statements:
OUTPUT 719;"sens:tcapture2:rang 4.01739e+19"
OUTPUT 719;"Tcapture2:Rang 9.85167e+19"
Query Syntax:
[SENSe:]TCAPture[1|2]:RANGe?
Return Format:
Real
Attribute Summary:
Synchronization Required:
Preset State:
SCPI Compliance:
Description:
The range must be specified before time data is loaded into the capture buffers from a controller.
This value (in Volts) is used as a reference to convert the data to integer format. To get full 32-bit
accuracy, the value sent should be the same as the largest data value to be loaded.
Note
Any data values larger than the range value will be clipped.
Related Commands:
To load data into the analyzer, use [SENS:]DATA.
To define the data format for the transfer, use FORM:DATA.
::= a real number (NRf data)
limits: 1.0E-20:1.0E20
::= MAX|MIN
0