To disinfect in situ, use a 0.1% sodium hypochloride solution (1000 ppm chlorine).
Do not use biological or phenolic-based cleaners. Removable, contaminated
covers should be laundered at 65°C for no less than 10 minutes or at 95°C for no
less than 3 minutes.
Thoroughly rinse and dry the cover before re-use or storage. Store in a cool, dry
area.
Do not mangle or iron.
Spin and tumble-dry at no more than 60°C.
3.3 Storage and Care
•
Store the mattress flat – do not fold.
•
Do not place heavy objects on the mattress surface when not in use.
•
Do not allow sharp objects to penetrate the cover or damage its
coating.
•
Keep the mattress clean.
•
Regularly inspect the interior and exterior of the mattress.
•
Do not store in damp conditions.
•
Store out of direct sunlight.
•
Store away from sources of ignition and sources of heat, such as,
radiators, open fires, heaters and electric heaters.
When removing packaging, use great care to ensure the cover is not damaged.
To aid storage and handling, store the mattress in a protective covering. When
removing packaging be careful to not damage or catch the mattress cover.
Always clean the mattress before putting back into service and inspect/audit the
mattress and cover.
3.4 Flammability
The Opera Serene meets the requirements of BS7177:2008 for medium hazard.
The mattress meets the requirements of The Furniture & Furnishings (Fire Safety)
Regulations 1988.