SUTO S011 Instruction And Operation Manual

Pressure transmitter

Advertisement

Quick Links

English
Instruction and Operation Manual
S011
Pressure Transmitter

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the S011 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for SUTO S011

  • Page 1 English Instruction and Operation Manual S011 Pressure Transmitter...
  • Page 2 The device is destined exclusively for the described application. SUTO offers no guarantee for the suitability for any other purpose. SUTO is also not liable for consequential damage resulting from the delivery, capability or use of this device.
  • Page 3: Table Of Contents

    5.4 Accuracy ................9 6 Dimensional Drawing..............9 7 Installation ................10 7.1 Installation Requirements............10 7.2 Installation Procedure............10 7.3 Electrical Connection.............11 8 Signal Output................12 8.1 Data Format.................12 8.2 Protocol Frame Analysis............12 8.3 Register................13 9 Configuration................14 10 Calibration................15 11 Maintenance................16 12 Disposal or Waste..............16 S011...
  • Page 4: Safety Instructions

    • Consider all regulations for electrical installations. • The system must be disconnected from any power supply during maintenance work. • Any electrical work on the system is only allowed by authorized qualified personal. S011...
  • Page 5 Measurement values can be affected by malfunction! The product must be installed properly and frequently maintained, otherwise it may lead to wrong measurement values, which can lead to wrong results. • Do not exceed the maximum operation temperature at the sensors tip. S011...
  • Page 6: Registered Trademarks

    • Avoid direct UV and solar radiation during storage. • For the storage the humidity has to be <90%, no condensation. 2 Registered Trademarks SUTO ® Registered trademark of SUTO iTEC MODBUS ® Registered trademark of the Modbus Organization, Hopkinton, USA S011...
  • Page 7: Application

    3 Application 3 Application The S011 Pressure Transmitter is designed to measure the pressure of compressed air and gases within the permissible operating parameter. Detailed parameters can be found in the technical data section. The S011 Pressure Transmitter can measure pressure values in MPa, kPa, or bar.
  • Page 8: Technical Data

    See dimensional drawing on the next page Screwing thread G 1/4” A (ISO 228/1) Electrical connection M12, 5 pins Stability ± 0.1% F.S. Vibration resistance 20 … 2000 Hz, 25g Weight 5.2 Electrical Data Power supply 24 VDC (12 ... 36 VDC) S011...
  • Page 9: Output Signals

    Response delay : 0 ms Inter-frame spacing : 7 char 5.4 Accuracy Accuracy ±0.25% F.S. S694 2559 / S694 2562 : S694 2563 ±0.5% F.S. : Repeatability ±0.1% F.S. S694 2559 / S694 2562 : S694 2563 ±0.25% F.S. : 6 Dimensional Drawing S011...
  • Page 10: Installation

    7 Installation Please make sure that all components listed below are included in your package. Qty Description Item No. S011 Pressure Transmitter S694 2559 (1.6 MPa (g)) S694 2562 (4.0 MPa (g)) S694 2563 (0.16 MPa (a)) M12 connector C219 0060...
  • Page 11: Electrical Connection

    Signal Colour Legend to pin assignment brown Not available white Negative supply voltage blue Positive supply voltage black Modbus data + grey Modbus data - ATTENTION! Do not screw the M12 plug using force. Otherwise, it may damage the connecting pins. S011...
  • Page 12: Signal Output

    The data received: 01 03 02 00 01 79 84 The function code of write-holding register is 06. The request frame format is as follows: Function Register Address Data Checksum Code Address 1 byte 1 byte 2 bytes 2 bytes 2 bytes S011...
  • Page 13: Register

    2: MPa 5: bar 6: psi 7: Pa * The value is related with the number of decimal places. For example, if the number of decimal places is 2 and the pressure value is 10023, the actual value is 100.23. S011...
  • Page 14: Configuration

    B. Add the pressure transmitter as a third-party device to the Modbus Master. C. Refer to the examples shown below to configure the pressure transmitter information including unit, resolution, value address (Modbus address), measure value type, function code, and display scaling. S011...
  • Page 15: Calibration

    However it is recommended to calibrate the instrument at least once per year. The calibration is excluded from the instruments warranty. For this please contact the manufacturer. S011...
  • Page 16: Maintenance

    The dispose can also be carried by the manufacturer of the product, for this please contact the manufacturer. SUTO iTEC GmbH SUTO iTEC (ASIA) Co., Ltd. Grißheimer Weg 21 Room 10, 6/F, Block B, Cambridge Plaza D-79423 Heitersheim 188 San Wan Road, Sheung Shui, N.T.

This manual is also suitable for:

S694 2559S694 2562S694 2563