Oxymax COS61D About this document About this document Warnings Structure of information Meaning This symbol alerts you to a dangerous situation. DANGER Failure to avoid the dangerous situation will result in a fatal or serious Causes (/consequences) injury. If necessary, Consequences of non-compliance (if applicable) ‣...
Basic safety instructions Oxymax COS61D Basic safety instructions Requirements for personnel • Installation, commissioning, operation and maintenance of the measuring system may be carried out only by specially trained technical personnel. • The technical personnel must be authorized by the plant operator to carry out the specified activities.
Oxymax COS61D Basic safety instructions 3. Do not operate damaged products, and protect them against unintentional operation. 4. Label damaged products as defective. During operation: ‣ If faults cannot be rectified: products must be taken out of service and protected against unintentional operation.
Device description, function Oxymax COS61D Device description, function Optical measuring principle Sensor structure Oxygen-sensitive molecules (markers) are integrated into the optically active layer (fluorescence layer). The fluorescence layer, an optical insulating layer and a cover layer are applied on top of one another on the carrier.
Oxymax COS61D Device description, function Sensor design A0042833 1 Sensor structure Sensor cable Threaded connection Sensor shaft Detector O-ring Emitter diode Protection guard Fluorescence cap The sensor consists of the following function units: • Sensor shaft • Sensor head with optics (emitter and detector) •...
Device description, function Oxymax COS61D Memosens technology Sensors with Memosens protocol have an integrated electronics unit that stores calibration data and other information. Once the sensor has been connected, the sensor data are transferred automatically to the transmitter and used to calculate the measured value.
Oxymax COS61D Incoming acceptance and product identification Incoming acceptance and product identification Incoming acceptance 1. Verify that the packaging is undamaged. Notify the supplier of any damage to the packaging. Keep the damaged packaging until the issue has been resolved.
5. Click on the product image in the popup window. A new window (Device Viewer) opens. All of the information relating to your device is displayed in this window as well as the product documentation. Manufacturer address Endress+Hauser Conducta GmbH+Co. KG Dieselstraße 24 D-70839 Gerlingen Scope of delivery Scope of delivery of sensor •...
Page 11
Oxymax COS61D Incoming acceptance and product identification Dieses Produkt erfüllt die Anforderungen der folgenden Normen: • ANSI/UL 61010-1, 3. Ed. • ANSI/UL 121201-2017 • ANSI/IEC 60529, Edition 2.2. 2013-08 Degrees of protection provided by enclosures (IP Code) Installation und Betrieb in explosionsgefährdeten Bereichen CL 1, DIV 2 Dieses nicht-funkende Gerät hat folgende spezifizierte Explosionsschutz-Daten:...
Installation Oxymax COS61D Installation Mounting conditions 5.1.1 Orientation 45° 135° 45° 0° 180° A0032281 2 Installation angle Recommended installation angle Possible installation angle Inadmissible installation angle The sensor must be installed at an angle of inclination in an assembly, holder or appropriate process connection.
Measuring system COS61D A complete measuring system consists of the following components at least: • Oxymax COS61D oxygen sensor with fixed cable (with ferrules or M12 plug depending on the version ordered) • Liquiline CM44x multi-channel transmitter • Assembly, e.g. flow assembly COA250, immersion assembly CYA112 or...
Page 14
Installation Oxymax COS61D 5.2.2 Mounting the cleaning unit or optional protection guard If the cleaning unit was not delivered as a pre-assembled unit or if an optionally ordered protection guard is used: A0042840 Unscrew the standard protection guard. Retain the standard protection guard for possible reuse at a later stage without the cleaning unit.
Page 15
Oxymax COS61D Installation 5.2.3 Installing at a measuring point Must be installed in a suitable assembly. WARNING Electrical voltage In the event of a fault, non-grounded metallic assemblies may be live and as such are not safe to touch! ‣...
Page 17
Oxymax COS61D Installation Universal holder and fixed immersion tube A0042859 6 Assembly holder with immersion tube Protective cover Transmitter Flexdip CYA112 immersion assembly Oxygen sensor Basin rim Assembly holder Flexdip CYH112 Basin rim mounting with immersion tube A0042860 7...
Page 18
Installation Oxymax COS61D Float The CYA112 float is for use in the case of large fluctuations in water level, for example in rivers or lakes. Cable run with strain relief and rain shield Fixing ring for rope and chains with terminal screw Eyelets Ø15, 3 x 120 °...
Oxymax COS61D Installation 5.3.3 Retractable assembly COA451 Detail A, 90 ° r Detail A A0030571 11 Permissible and impermissible sensor installation positions with retractable assembly COA451 Ascending pipe, best position Horizontal pipe, sensor top down, impermissible due to air cushion or foam bubble forming Horizontal pipe, lateral installation with permissible installation angle (acc.
Electrical connection Oxymax COS61D Electrical connection WARNING Device is live! Incorrect connection may result in injury or death! ‣ The electrical connection may be performed only by an electrical technician. ‣ The electrical technician must have read and understood these Operating Instructions and must follow the instructions contained therein.
Oxymax COS61D Calibration and adjustment Calibration and adjustment Types of calibration The following types of calibration are possible: • Zero point • Single-point calibration in nitrogen or COY8 zero-point gel • Numeric input • Slope • Air 100% rh (air, water vapor-saturated) •...
Calibration and adjustment Oxymax COS61D 6. Is the measured value display on the transmitter stable: Perform the calibration in accordance with the Operating Instructions for the transmitter. Pay particular attention to the software settings for the stability criteria for calibration and for the ambient pressure.
Commissioning Oxymax COS61D Commissioning Function check Prior to initial commissioning, ensure that: • The sensor is correctly installed • The electrical connection is correct If using an assembly with automatic cleaning function: ‣ Check that the cleaning medium (water or air, for example) is connected correctly.
Oxymax COS61D Troubleshooting Troubleshooting Troubleshooting instructions ‣ If one of the following problems is present: Check the measuring system in the order shown. Problem Testing Remedial action ‣ Nothing displayed, no Power supplied to the transmitter? Establish the power supply.
Maintenance Oxymax COS61D Maintenance Take all the necessary precautions in time to ensure the operational safety and reliability of the entire measuring system. NOTICE Effects on process and process control! ‣ When carrying out any work on the system, bear in mind any potential impact this could have on the process control system and the process itself.
Oxymax COS61D Maintenance Clean the sensor: • Before every calibration • At regular intervals during operation as necessary • Before returning it for repairs Type of contamination Cleaning Salt deposits Immerse the sensor in drinking water or in 1-5 % hydrochloric acid (for a few minutes).
Maintenance Oxymax COS61D reaches a specific value. The sensor is still able to measure at this stage. However, it is advisable to change the cap as quickly as possible. Removing the old fluorescence cap 1. Activate the hold function at the transmitter.
Oxymax COS61D Accessories Accessories The following are the most important accessories available at the time this documentation was issued. ‣ For accessories not listed here, please contact your Service or Sales Center. 11.1 Assemblies (selection) Flexdip CYA112 • Immersion assembly for water and wastewater •...
Accessories Oxymax COS61D 11.4 Zero-point gel COY8 Zero-point gel for oxygen and disinfection sensors • Oxygen-free and chlorine-free gel for the verification, zero point calibration and adjustment of oxygen and disinfection measuring points • Product Configurator on the product page: www.endress.com/coy8...
Oxymax COS61D Accessories 11.8 Transmitter Liquiline CM44 • Modular multi-channel transmitter for hazardous and non-hazardous areas • Hart®, PROFIBUS, Modbus or EtherNet/IP possible • Order according to product structure Technical Information TI00444C Endress+Hauser...
The product must be returned if repairs or a factory calibration are required, or if the wrong product was ordered or delivered. As an ISO-certified company and also due to legal regulations, Endress+Hauser is obliged to follow certain procedures when handling any returned products that have been in contact with medium.
Technical data Oxymax COS61D 13.4 Process Process temperature –5 to +60 ˚C (20 to 140 ˚F) Process pressure Ambient pressure 1 to 10 bar (14.5 to 145 psi) abs. 13.5 Mechanical construction Dimensions NPT ¾" Ø40 (1.57) A0042897 A0042896 14 With optional cleaning unit ...
Page 35
Oxymax COS61D Technical data Optional cleaning unit 19 (0.75) 28.5 (1.12) 61 (2.40) A0043286 15 Dimensions in mm (inch) Compressed air cleaning COS61/61D OD 6/8 mm (additional information → 30 Compressed air cleaning COS61/61D OD 6.35 mm (¼") (additional information → 30...
Page 36
Technical data Oxymax COS61D Interface Memosens protocol Endress+Hauser...