11.2 Power failure alarm

If at any time the mains power should be removed from the power unit or the power cable is
unplugged without turning the power unit OFF, an audible alarm will be heard and the Power Failure
indicator will illuminate. The audible alarm may be cancelled by pressing the Alarm Mute button.
Note: When the power unit has not been used for more than 3 months, it might
require 60 minutes operating time (or more) for the Alarm to function correctly.

11.3 Service (alternating failure alarm)

Should your system develop a fault condition whilst in use, an audible alarm will be heard and the
Service indicator will illuminate. Please refer to Table 1 for error codes and call Ultimate Healthcare
or your local distributor.

11.4 Annual Service Indicator

A dedicated Annual Service indicator on the control panel will illuminate and stay ON to alert nursing
staff of the need for the system to be professionally serviced.
The Annual Service indicator will illuminate after 365 days of use and the indicator light can only be
extinguished by a professional service technician once the required service has been conducted.
Table 1: Alert/Error Code Reference Table
Priority
High
Warning
Indicator
Code
Low
0
N/A
Power
1
Failure
2
ALL LED
3
N/A
4
Auto-Firm
5
Auto-Firm
6
Static
Power
7
N/A
Failure
8
Pressure
9
Pressure
Issue: 4 - 16/08/18
Audible
LED
Output Mode
N/A
ONCE
ONCE
ONCE
N/A
ONCE
ONCE
ONCE
ONCE
REPEAT
(Cycle 4 sec.)
Low
REPEAT
(Cycle 4 sec)
Low
REPEAT
(Cycle 4 sec)
Condition of
Warning Description
Output
Not in System
Key Tone
Shutdown
Power-Off
System Shutdown
Operate or
Power-On
Standby
Operate or
State / Mode
Standby
Switching
Mattress Inflation
Operate
Completion
Operate
Auto-Firm Completion
Operate
Static Completion
Power-Off
Power Failure Alarm
Operate or
Power-On Inflation
Standby
Failure Alarm
Operate or
Auto-Firm
Standby
Failure Alarm
Remarks
Key Tone from
Functional Button
Shutdown
All Indicators
Light On
No Display
Inflation Ended
Auto-Firm Ended
Static Ended
No Display
Inflate Failure
Auto-Firm Failure
Page 16 of 29