National Instruments SCXI-1194 Getting Started Manual page 9

Table of Contents

Advertisement

1. Chassis Power Switch
2. Chassis Address Switches
1.
Power off and unplug the chassis.
2.
Install the switch controller, including any adapter kit(s), as described in the switch
controller documentation.
Note Ensure the chassis is still powered off before continuing to install your
switch module.
3.
Set the chassis address switches to the OFF position. If you are using multiple SCXI
chassis, ensure each chassis has a unique binary address.
Note Earlier chassis use jumpers inside the chassis front panel instead of
chassis address switches. Earlier chassis also differ in fuses and AC power
selection. Refer to the documentation specific to your chassis for more
information.
4.
Confirm the correct power settings of the chassis (100, 120, 220, or 240 VAC).
5.
Remove the filler panel from an unused SCXI slot.
6.
Touch any metal part of the chassis to discharge static electricity.
7.
Insert the switch module into the SCXI slot.
8.
Screw the switch front panel to the chassis front panel mounting rail.
8 | NI Switches Getting Started Guide | ni.com
Figure 2. Chassis Preparation
Front
P O
W E
R
1
R E
S E
T
5
2
4
3
2
1
O N
RESET
5
4
3
2
1
ON
Back
~
c
V a
1 2 0
S
C
X I
3. Chassis Power Settings
4. Power Cord Connector
3
4

Advertisement

Table of Contents
loading