The RD3 serial interface can automatically select the serial input protocol from the serial output
of an indicator in one of the following serial formats:
SMA, Cardinal SB400, Numbers, Rice Lake IQ355, ANDFV,
WI110, Toledo Short, and Toledo Long
The interface will automatically detect baud rates between 2400 and 19200 baud with the
following settings:
Baud rate:
Character format:
Electrical Power
The RD3 display has been designed to operate from 100 to 240V AC at 47/63 Hz.
CAUTION! To avoid electrical hazards and possible damage to the display,
DO NOT, under any circumstance, cut, remove, alter, or in any way bypass the
power cord-grounding prong.
• The power for the display should be on a separate circuit from the distribution panel. This
circuit should be dedicated to the exclusive use of the display.
• The wiring should conform to national and local electrical codes and ordinances and should
be approved by the local inspector to ensure compliance.
• For permanently connected equipment, a readily accessible disconnect device must be
provided external to the equipment.
• For pluggable equipment, the socket outlet shall be installed near the equipment and shall
be easily accessible
• It is the responsibility of the customer to have a qualified electrician install the proper
service disconnect that conforms to national electrical codes and local codes and
ordinances.
Electrical Noise Interference
To prevent electrical noise interference, make certain all other wall outlets for use with air
conditioning and heating equipment, lighting, or other equipment with heavily inductive loads,
such as welders, motors, and solenoids are on circuits separate from the display. Many of
these disturbances originate within the building itself and can seriously affect the operation of
the display. These sources of disturbances must be identified and steps must be taken to
prevent possible adverse effects on the display. Examples of available alternatives include
isolation transformers, power regulators, uninterruptible power supplies, or simple line filters.
2
8200-M063-O1 Rev C  RD3 Installation and Technical Manual

AUTO-FORMAT FEATURE

2400 to 19200 (7 or 8 Data Bits)
all standard ASCII characters

SITE PREPARATION