Thermo Scientific Forma 8602 Operating And Maintenance Manual page 45

-86c ult freezer
Table of Contents

Advertisement

Section 6
Factory Options
Injection Assembly, Vent
Stack and Solenoid (cont.)
Figure 6-1. Injection Assembly
6-2
Model 8600 Series
6. Install the vent stack through the opening and secure it to the top of
the freezer, using screws.
7. Go to the interior and seal around the end of the vent stack with
Permagum.
8. Install the transfer hose connecting one end to the injection assembly,
the other end to the solenoid valve. Install the solenoid valve to the
supply source. The solenoid mounting bracket is not required and may
be discarded.
Caution When selecting a CO
siphon tube.
supply cylinder, it must be equipped with a
2
Figure 6-2. Vent Stack