National 215-RLED Instructions Manual page 7

Compound microscope
Table of Contents

Advertisement

Your microscope includes 3 rechargeable AA nickel metal hydride batteries. These may be recharged up to 500
times, but if you observe that a recharge is providing significantly less than 40 hours of operation. It is probably
time to replace to batteries.
IMPORTANT WARNING: DO NOT USE REGULAR ALKALINE BATTERIES IN THIS MICROSCOPE. ANY
ATTEMPT TO RECHARGE ALKALINE TYPE BATTERIES COULD RESULT IN BATTERIES EXPLODING.
Gently lay microscope on its side or back. Observe small door in center of microscope bottom base. Loosen
slotted screw on door and open. Battery case is mounted on inside of door. Using small Phillips screwdriver,
carefully remove Phillips screw that holds battery case together. Slide lid of case straight out to remove and
expose batteries. Remove all 3 batteries and replace with new rechargeable AA nickel metal hydride batteries,
making certain to insert with correct polarity according to markings on battery holder. Replace lid, close and
secure door.
Follow instructions on new battery packaging to determine if they are already charged, or if they should be charged
before initial use. If recharging is required, following directions in "3.a" above.
c. Replacing LED element:
An LED "bulb" will last up to 50,000 hours, so you don't have to do this exercise very often.
To open the illuminator field lens housing, use hex wrench supplied with your microscope to loosen the two hex
screws on side of lens housing. Remove lens housing to expose LED "bulb." Remove bulb by grasping the
plastic base of bulb and gently pulling straight up. Insert new LED "bulb", replace lens housing and tighten hex
screws to secure lens housing in place.