Siemens SIPROCESS GA700 Compact Operating Instructions page 88

Continuous gas analytics
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Technical data
7.1 General technical specifications
Electrical inputs and outputs
Digital outputs (DO)
Digital inputs (DI)
Ethernet interface (rear)
Table 7-2
Rack-mounted housing: Option modules
Option modules
Option module 1.1
Option module 2.1
Option module 2.2 I/O
Option module 2.2 Profibus DP
Table 7-3
Rack-mounted housing: Climatic conditions
Climatic conditions
Permissible operating altitude
Permissible ambient temperature
for OXYMAT 7 and CALOMAT 7
Permissible ambient temperature
during operation for ULTRAMAT 7
Storage and transport
Ambient humidity (relative humid‐
ity) during storage, transport or op‐
eration
Pollution degree
1.) Reduction of permissible temperature range in case of operation in combination with Ultramat 7.
88
8, with changeover contacts, parameters can be freely assigned, e.g. for measuring range
identification; max. load: 24 V AC/DC/1.7 A (total load for all 8 relay outputs in continuous
operation max. 160 W), isolated, non-sparking
8, designed for 24 V, isolated, freely parameterizable, e.g. for autoranging
Ethernet RJ 45, 100 Mbps
12 digital outputs, with changeover contacts, max. load: 24 V AC/DC / 1.7 A
(total load for all 12 relay outputs in continuous operation max. 244 W), isolated, non-
sparking
8 digital inputs, designed for 24 V, isolated, freely configurable
6 analog outputs 0/4 ... 20 mA, designed for 24 V, isolated; load:  R
6 digital outputs, load rating: 24 V AC/DC/1.7 A (total load for all 6 relay outputs in
continuous operation max. 122 W), isolated, non-sparking
6 analog outputs 0/4 to 20 mA, designed for 24 V, isolated; max. load 750 Ω
4 analog inputs 0/4 to 20 mA, designed for 24 V, non-isolated, internal resistance
100 Ω
4 digital inputs, designed for 24 V, isolated
no additional inputs and outputs
Profibus communication
3 000 m above sea level sea level
0 ... 50 °C (32 ... 122 °F)
1)
5 °C ... +45 °C (41 ... +133 °F)
- 30 ... +70 °C (-22 ... 158 °F)
< 90%
No cooling below dew point during storage and operation
2
SIPROCESS GA700 rack and wall mounted devices
Compact Operating Instructions, 06/2023, A5E35134047-AB
 ≤ 750 Ω
L

Advertisement

Table of Contents
loading