unit has reached the optimum conditions for ignition.
– If a bulb fails to ignite, 150 seconds after the OPERATE switch is turned ON:
• the LASER ON LED will be OFF.
• the LAMP FAULT LED will be ON.
• the LAMP ON LED will remain OFF.
This is very unusual. However, if this occurs, turn the OPERATE switch to the OFF position ("down" position)
and restart this procedure at Step 1. If this issue occurs multiple times, see
contact Energetiq's Service department.
5.2 Stopping the System
1. To turn the EQ-99X system OFF, simply turn the OPERATE switch OFF by placing the EQ-99-RC Remote Control
switch in the "down" position.
2. If the system will not be used for some time, turn the 12VDC supply off.
Note: To minimize wear on the ignition components, it is best to avoid frequently starting and stopping the
EQ-99X system. It is recommended to run the system continuously rather than turn the system off and on
several times in one day.
EQ-99X Operation & Maintenance Manual
Chapter 5. Operating the System
Chapter 7.
Troubleshooting or
35 of 45