Hamamatsu ENERGETIQ LDLS EQ-99X Operation And Maintenance Manual page 13

Laser-driven light source
Table of Contents

Advertisement

   
LED Label
LAMP MODULE FAULT
 
B. Input/Output (I/O) Port
The I/O port is used to connect a remote control to the EQ-99X system. You can either use the EQ-99-RC Remote
Control provided with the system, or another remote control. See 4.4.2 Installing Alternative Remote
assignments and functions.
 
C. Power Port
This is a latching port for 12VDC power. Power can be provided using the included 12VDC power supply. Alternatively,
the EQ-99X
system can be powered directly from a customer-provided 12VDC power supply. See 4.4.1 Installing
Alternative Power
Supply for detailed information.
D. Lamp Head Signal Port
This port is used to provide various power and control signals to/from the lamp head. No other connector or cable
may be used with the EQ-99X system other than the one supplied.
E. Laser Fiber Optic Cable
The laser light is delivered from the power supply controller to the lamp head via a fiber optic cable with armored 
protection. The fiber is permanently attached to the power supply controller and connects to the lamp head via an
SMA-type connector.
 
It is critical that this armored fiber be treated with care and inspected for any abnormalities prior to operation. Avoid
sharp bends which can permanently damage the fiber. The minimum bend radius is 30 mm (1.18 in). Avoid crimping
or compressing the fiber. See Chapter 4. Installation for more information on installation and Chapter 6.
Maintenance for more information on fiber end-face cleaning.
 
  EQ-99X Operation & Maintenance Manual
Meaning (when lit)
One of the following has occurred in the lamp head:
1. Control cable not connected properly
2. Lamp head internal temperature too high
3. Laser fiber not correctly connected
4. Lamp bulb not correctly installed
5. Ignition failure
Chapter 2. System Description  
Control for pin 