Filter Change - ULT LAS260.1 Operating Instructions Manual

Extraction and filtration unit
Table of Contents

Advertisement

Filter change

NOTICE
Page 30
Maintenance of the device is limited to the inspection of filter
saturation and replacement of the filter.
The LED status ring indicates the status of the particulate filters
(pre-filter and main filter). Filter saturation can also be reported
via the M12 interface to a higher-level control.
Color code
Green
• Particulate filters are OK.
Orange
• Particulate filters are heavily
• The particulate filters need to be
• The particulate filters are completely
Red
• To start with, only replace the pre-
The saturation of the adsorption filter is not indicated via the
LED status ring.
After a certain period of time, depending on the level of
pollutants/impurities present the activated carbon will become
saturated. The time until saturation occurs needs to be
determined, and the filter should then be regularly replaced
whenever this time limit is reached. If the gases that are
supposed to be separated out are noticeably present in the
outlet air stream, change the adsorption filter immediately.
Maintenance and servicing
Status
saturated.
replaced soon.
saturated and must be replaced
immediately.
filter. If the situation does not
improve after replacement, then
change the main filter.
LAS 260.1 – Date 07/2023

Advertisement

Table of Contents
loading

Table of Contents