Shutdown - ULT LAS260.1 Operating Instructions Manual

Extraction and filtration unit
Table of Contents

Advertisement

Remote operation
NOTICE

Shutdown

Page 28
In remote mode, the Extraction and Filtration System
LAS 260.1 can be controlled by a higher-level machine via the
communication interface M12.
The higher-level machine can then switch on the LAS 260.1
and set it into standby mode.
Prerequisite for remote operation:
The higher-level machine is connected via the
-
communication interface M12 to the LAS 260.1.
S2 switch is in the position "0"
-
The operator can still adjust the suction power by hand on the
potentiometer and switch on the ecoflow CS
(see section ecoflow CS
If no higher-level machine is connected, the S2 switch must be
in the position "I."
The time interval between switching on and switching off
should be at least 5 minutes. Otherwise, there is a risk that the
blower might be damaged.
When shutting down the LAS 260.1, always follow the safety
instructions in these operating instructions.
-
Switch off the unit.
-
Disconnect the electrical connections of the LAS 260.1.
Shutdown
®
®
).
LAS 260.1 – Date 07/2023
via the S3 switch