OPERATION

  • The meter is supplied complete with a 9V battery. Slide off the battery compartment cover on the back of the meter. Install the battery while paying attention to its polarity.
  • Always remove the electrode protective cap before taking any measurement. If the electrode has been left dry, soak the tip (bottom 2.5 cm) in rinse solution (M10000B) for a few minutes to reactivate it.
  • Connect the pH electrode to the BNC socket on the top of the meter.
  • Make sure that the meter has been calibrated before taking any measurements (see Calibration Procedure)
  • Immerse the tip (2.5 cm) of the pH electrode into the sample.
  • Turn the instrument on by pressing the ON/OFF key.
  • Allow the reading to stabilize before taking measurements.
  • After measurements, switch the meter off and store the electrode with a few drops of storage solution in the protective cap.

CALIBRATION PROCEDURE

CALIBRATION PROCEDURE

  • Remove the protective cap from the electrode.
  • Immerse the electrode tip into a new sachet of pH7 calibration solution and allow the reading to stabilize.
  • Adjust the pH7 (OFFSET) calibration knob to display "7.0 pH".
  • Open a new sachet of pH4 calibration solution and use a small quantity to rinse the electrode.
  • Immerse the electrode tip inside the pH4 sachet and allow the reading to stabilize.
  • Adjust the pH4/pH10 (SLOPE) calibration knob to read "4.0 pH".
  • The calibration is now complete and the meter is ready to take measurements.
  • It is suggested to recalibrate the meter at least once a month, after a prolonged stocking time and after electrode replacement.

BATTERY REPLACEMENT

When the battery becomes weak the meter will display "V".

When the low battery indicator appears, the battery has only about 50 hours of working time left. A low battery will result in unreliable measurements. It is recommended to replace immediately the battery.

Battery replacement must only take place in a non-hazardous area using an alkaline 9V battery.

Turn the meter off, slide off the battery compartment cover at the rear of the meter and replace the 9V battery with a new one. Make sure the battery contacts are tight and secure, seat the battery in its compartment and insert the cover again.

SPECIFICATIONS

RANGE 0.0 to 14.0 pH
RESOLUTION 0.1 pH
ACCURACY (@25°C) ±0.2 pH
pH ELECTRODE MA911B/1 (included)
ENVIRONMENT 0 to 50°C, 95% RH max.
BATTERY TYPE 1 x 9V alkaline (included)
BATTERY LIFE approximately 300 hours of use
DIMENSIONS 145 x 80 x 40 mm
WEIGHT 220 g (with battery) meter only

OPTIONAL ACCESSORIES

M10004B pH 4.01 buffer solution, 20 ml sachet (25 pcs)
M10007B pH 7.01 buffer solution, 20 ml sachet (25 pcs)
M10010B pH 10.01 buffer solution, 20 ml sachet (25 pcs)
M10000B Electrode rinse solution, 20 ml sachet (25 pcs)
MA9015 Electrode storage solution, 220 ml bottle
MA911B/1 pH-electrode with BNC connector and 1 m cable
MA950 Portable meter wall fixing kit

Documents / Resources

Download manual

Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.

Download Milwaukee Smart PH Meter, SM100 - Smart PH Meter Manual

Advertisement

Need help?

Need help?

Do you have a question about the Smart PH Meter and is the answer not in the manual?

Questions and answers

Table of Contents