Milwaukee Smart PH Meter, SM100 - Smart PH Meter Manual
![]()
OPERATION
![]()
- The meter is supplied complete with a 9V battery. Slide off the battery compartment cover on the back of the meter. Install the battery while paying attention to its polarity.
- Always remove the electrode protective cap before taking any measurement. If the electrode has been left dry, soak the tip (bottom 2.5 cm) in rinse solution (M10000B) for a few minutes to reactivate it.
- Connect the pH electrode to the BNC socket on the top of the meter.
- Make sure that the meter has been calibrated before taking any measurements (see Calibration Procedure)
- Immerse the tip (2.5 cm) of the pH electrode into the sample.
- Turn the instrument on by pressing the ON/OFF key.
- Allow the reading to stabilize before taking measurements.
- After measurements, switch the meter off and store the electrode with a few drops of storage solution in the protective cap.
CALIBRATION PROCEDURE
![]()
![Milwaukee - Smart PH Meter - CALIBRATION PROCEDURE CALIBRATION PROCEDURE]()
- Remove the protective cap from the electrode.
- Immerse the electrode tip into a new sachet of pH7 calibration solution and allow the reading to stabilize.
- Adjust the pH7 (OFFSET) calibration knob to display "7.0 pH".
- Open a new sachet of pH4 calibration solution and use a small quantity to rinse the electrode.
- Immerse the electrode tip inside the pH4 sachet and allow the reading to stabilize.
- Adjust the pH4/pH10 (SLOPE) calibration knob to read "4.0 pH".
- The calibration is now complete and the meter is ready to take measurements.
- It is suggested to recalibrate the meter at least once a month, after a prolonged stocking time and after electrode replacement.
BATTERY REPLACEMENT
![]()
When the battery becomes weak the meter will display "V".
When the low battery indicator appears, the battery has only about 50 hours of working time left. A low battery will result in unreliable measurements. It is recommended to replace immediately the battery.
Battery replacement must only take place in a non-hazardous area using an alkaline 9V battery.
Turn the meter off, slide off the battery compartment cover at the rear of the meter and replace the 9V battery with a new one. Make sure the battery contacts are tight and secure, seat the battery in its compartment and insert the cover again.
SPECIFICATIONS
RANGE | 0.0 to 14.0 pH |
RESOLUTION | 0.1 pH |
ACCURACY (@25°C) | ±0.2 pH |
pH ELECTRODE | MA911B/1 (included) |
ENVIRONMENT | 0 to 50°C, 95% RH max. |
BATTERY TYPE | 1 x 9V alkaline (included) |
BATTERY LIFE | approximately 300 hours of use |
DIMENSIONS | 145 x 80 x 40 mm |
WEIGHT | 220 g (with battery) meter only |
OPTIONAL ACCESSORIES
M10004B | pH 4.01 buffer solution, 20 ml sachet (25 pcs) |
M10007B | pH 7.01 buffer solution, 20 ml sachet (25 pcs) |
M10010B | pH 10.01 buffer solution, 20 ml sachet (25 pcs) |
M10000B | Electrode rinse solution, 20 ml sachet (25 pcs) |
MA9015 | Electrode storage solution, 220 ml bottle |
MA911B/1 | pH-electrode with BNC connector and 1 m cable |
MA950 | Portable meter wall fixing kit |
![]()
Documents / ResourcesDownload manual
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Download Milwaukee Smart PH Meter, SM100 - Smart PH Meter Manual
Need help?
Do you have a question about the Smart PH Meter and is the answer not in the manual?
Questions and answers