WIKA mensor CPG2500 Operating Instructions Manual

WIKA mensor CPG2500 Operating Instructions Manual

Vacuum calibrator

Advertisement

Quick Links

Operating Instructions
Vacuum Calibrator
CPG2500
Vacuum Calibrator CPG2500
PN 0018906001• 10/2019

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the mensor CPG2500 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for WIKA mensor CPG2500

  • Page 1 Operating Instructions Vacuum Calibrator CPG2500 Vacuum Calibrator CPG2500 PN 0018906001• 10/2019...
  • Page 2 Vacuum Calibrator CPG2500 This Warning symbol indicates that danger of injury for persons and the environment and/or considerable damage (mortal danger, danger of injury) will occur if the respective safety precautions are not taken. Warning This Caution symbol indicates danger for the system and material if the respective safety precautions are not taken.
  • Page 3: Table Of Contents

    Vacuum Calibrator CPG2500 Table of Contents 1. General Information 1.1 Warranty 1.2 Important Notice 1.3 FCC Radio Frequency Emission Notice 1.4 Trademarks and Copyrights 1.5 Software License Agreement 1.6 Mensor Service Plus 1.6.1 After the Warranty 1.6.2 Calibration Services 1.6.3 Certifications and Accreditations 1.7 Packaging for Shipment 2.
  • Page 4 Vacuum Calibrator CPG2500 6.1.2 Display Screen Features 6.2 Initial Setup 6.2.1 Contact and Version Information Application 6.2.2 Language Selection 6.3 Application Selection and parameter inputs 6.4 Applications: 6.4.1 Measure Application 6.4.1.1 Auxiliary Displays 6.4.1.2 Zero Button 6.4.2 Settings Application 6.4.2.1 Languages 6.4.2.2 Brightness...
  • Page 5 Vacuum Calibrator CPG2500 7.5 Mensor Command Set 7.6 Command and Query Format 7.7 Command Set Definitions 7.8 Output Formats 7.9 CPG2500 Commands and Queries 7.9.1 Units Command Syntax for Measurement Units 7.9.2 CPG2500 Error Codes 7.9.3 GPIB Capability Codes 7.9.4 Interface Functions 8.
  • Page 6: General Information

    Mensor has made a concerted effort to provide complete and current information for the proper use of the equipment. If there are questions regarding this manual or the proper use of the equipment, contact Mensor at: Mensor WIKA Alexander Wiegand SE & Co. KG 201 Barnes Drive Alexander-Wiegand-Straße 30 San Marcos, Tx 78666 D-63911 Klingenberg / Germany Tel: 1.512.396.4200...
  • Page 7: Trademarks And Copyrights

    Strictly observe the following when shipping the instrument: All instruments delivered to WIKA or Mensor must be free from any kind of hazardous substances (acids, bases, solutions, etc.) and must be cleaned before being returned.
  • Page 8 Vacuum Calibrator CPG2500 WARNING! Physical injuries and damage to property and the environment through residual media Residual media in the dismounted instrument can result in a risk to persons, the environment and equipment. With hazardous substances, include the material safety data sheet for the corresponding medium. When returning the instrument, use the original packaging or a suitable transport packaging.
  • Page 9: Safety

    Vacuum Calibrator CPG2500 Safety User Responsibilities To ensure safety, the user must make sure that: • The system is used properly, no dangerous media are used and that all technical specifications are observed. • The system is operated in perfect operating condition. •...
  • Page 10: Warnings And Caution Notices

    Vacuum Calibrator CPG2500 Warnings and Caution Notices WARNING: NOT EXPLOSION PROOF! Installation of this instrument in an area requiring devices rated as intrinsically safe is not recommended. Warning WARNING: POSSIBLE INJURY! The tubing, valves, and other apparatus attached to the gauge must be adequate for the maximum pressure which will be applied, otherwise physical injury to the operator or bystanders is possible.
  • Page 11: General Description

    Vacuum Calibrator CPG2500 General Description The CPG2500 Vacuum Calibrator is a precision vacuum indicator and calibrator designed to test and calibrate absolute pressure transducers and indicators. The CPG2500 has a precision vacuum transducer capable of measuring vacuum pressure with 1 mTorr accuracy. Figure 3.1 - CPG2500 Vacuum Calibrator Features Here is a short list of significant features designed into the CPG2500:...
  • Page 12: Turning On

    Vacuum Calibrator CPG2500 Turning On You can confirm that your CPG2500 is operational right now. Apply power to the power connector on the rear of the instrument with the included power adapter, remove any plastic plugs from the rear panel pressure ports, and press the power switch to ON.
  • Page 13: Display

    Vacuum Calibrator CPG2500 Display The display is made up of two sections. In the main screen (“Measure Application”), the left two thirds shows the internal and DUT channels, along with the pressure reading, units, mode (absolute), channel label, an auto zero button (only on DUT channel) and any auxiliary indications that have been chosen. The right one third of the screen has Application Icons (“Apps”) for setting viewing options, sensor settings, communication settings plus a “Page Down”...
  • Page 14: Specifications

    Vacuum Calibrator CPG2500 Specifications Accuracy specifications presented herein are obtained by comparison with primary standards traceable to a national metrology institute or recognized international standard organization. These specifications are obtained in accordance with the ISO Guide to the Expression of Uncertainty in Measurement (GUM). The calibration program at Mensor is accredited by the American Association of Laboratory Accreditation (A2LA) as complying with both the ISO/IEC 17025:2017 and the ANSI/NCSL Z540-1-1994 standards.
  • Page 15 Vacuum Calibrator CPG2500 Switching power supply Output: 12 VDC, 1.7 A (includes 4 region specific plugs adapters) Permissible ambient conditions Storage temperature range 0 to 70 deg C Operating environment 0 ... 95 % RH (relative humidity, non-condensing) Operating temperature range 15 ...
  • Page 16: Installation

    Vacuum Calibrator CPG2500 Installation Unpacking the Instrument In addition to functional testing, each unit is inspected for appearance prior to leaving the factory. Upon receipt, please examine the instrument for shipping damage. Report any apparent damage to the carrier immediately. In addition to this manual you should have: •...
  • Page 17: Dimensions (Inches)

    Vacuum Calibrator CPG2500 Dimensions (inches) Desk top case Front view Side view Rear Panel IEEE-488 DUT/ Remote Transducer connection RS-232 USB (device) USB (host) Ethernet port Power Measure port (vacuum calibrator) 1/4” NPT female Operating Instructions - CPG2500 Vacuum Calibrator...
  • Page 18: Mounting

    Vacuum Calibrator CPG2500 Mounting The instrument is set up on a desk top. The vacuum transducer built in the CPG2500 is sensitive to vibration, to further assure stability and accuracy, avoid mounting the instrument on surfaces subject to excessive motor or machinery vibration. Rear Panel The measure port for the vacuum calibrator is located at the rear panel of the CPG2500.
  • Page 19: Remote Communication Connections

    Vacuum Calibrator CPG2500 Remote Communication Connections See Section 7, Remote Operation, for connections and commands for operation over IEEE-488, Ethernet, USB or RS-232 ports. 5.5.1 Remote Transducer Port The Remote Transducer Port is used to connect the Device Under Test (DUT) to the CPG2500 Vacuum Calibrator.
  • Page 20 Vacuum Calibrator CPG2500 Figure 5.7-A T-fitting on Vacuum Pump 8 mm pressure tube Needle valve to connecting DUT regulate vacu- pressure port to the um supply vacuum pump T shaped connection to source vacuum supply to DUT and CPG2500 Vacuum pump 8 mm pressure CPG2500 tube con-...
  • Page 21: Local Operation And Setup

    Vacuum Calibrator CPG2500 Local Operation and Setup General Operation This section describes the procedures for operating the CPG2500 from the front panel. Instructions for operating the device remotely from an external computer are covered in Section 7, Remote Operation. By following the procedures provided in these two sections and Section 10, Calibration, you can expect your CPG2500 to deliver maximum accuracy and dependability for many years of useful service.
  • Page 22 Vacuum Calibrator CPG2500 Initial Setup Section 6.2.1 and 6.2.2 are provided first so that the operator can initially check the information screen to verify the installed components and to change the language if needed. 6.2.1 Contact and Version Information Application Press this application button to display Mensor contact, instrument and software version information.
  • Page 23 Vacuum Calibrator CPG2500 Application Selection and parameter inputs The application selection area on the right one third of the screen (see Figure 6.1.2 - Display Screen Features) is the area where setup, information and calibration Apps can be chosen. A second and third page of application selections can be accessed by pressing the page down button [ ].
  • Page 24 Vacuum Calibrator CPG2500 Applications: 6.4.1 Measure Application The Measure App is the normal operation screen referred to in previous instruments as the ‘main screen’. This application is different from the others in that it is not use to setup the configuration but is used to monitor the pressure applied to the installed transducers.
  • Page 25 Vacuum Calibrator CPG2500 6.4.1.1 Auxiliary Displays The screen in figure 6.4.1.1 shows all of the possible auxiliary display items that can be included in the Measure App. The auxiliary display item includes “Alternate Units”, “Peak” and “Rate”. Additional “Zero” calibration function appears on the DUT (remote) channel. These auxiliary features will appear in the Measure App when selected from the Transducer App (section 6.4.3.3).
  • Page 26 Vacuum Calibrator CPG2500 Figure 6.4.1.3-A - Zero Grayed. Vacuum Too High Figure 6.4.1.3-B - Zeroing Figure 6.4.1.3-C - Zero Successful Operating Instructions - CPG2500 Vacuum Calibrator...
  • Page 27 Vacuum Calibrator CPG2500 6.4.2 Settings Application The Settings App is used to set up general settings for the display. Settings parameters include Language, Brightness, Volume, User 1 base units, User 1 multiplier, User 2 base units, User 2 multiplier, Barometer units, and Configuration. Figure 6.4.2 shows these parameters as indicated when the Settings App has been chosen.
  • Page 28 Vacuum Calibrator CPG2500 6.4.2.2 Brightness The Brightness setting provides a sliding scale to increment the screen brightness in all screens. Sliding your finger along the bar graph or touching anywhere in the bar graph will change the brightness of the screen.
  • Page 29 Vacuum Calibrator CPG2500 6.4.2.4 User base units / Base units multiplier When choosing a unit of measure from the Measure Application (main screen), standard units can be chosen in addition to two user defined units. User units 1 and 2 are defined in the Settings App using “User 1 base units”, “User 1 multiplier”...
  • Page 30 Vacuum Calibrator CPG2500 Figure 6.4.2.6 - Configuration The instrument default configuration can be activated simply by pressing the “Default” Button. Operating Instructions - CPG2500 Vacuum Calibrator...
  • Page 31 Vacuum Calibrator CPG2500 6.4.3 Transducer Application For the internal vacuum channel, the Transducer Application provides a way to set the filter for the reading to reduce fluctuations due to electrical noise, and to set the resolution of the reading. In addition, the auxiliary display functions can be specified here. Figure 6.4.3 - Transducer Application 6.4.3.1 Transducer Filter The Filter is an electronic filter to smooth out the pressure readings.
  • Page 32 Vacuum Calibrator CPG2500 6.4.3.2 Transducer Resolution The resolution of each transducer channel can be set in the Transducer Application using the resolution parameter. The resolution can be set to 4, 5, or 6 digits. Figure 6.4.3.2 - Transducer Resolution Operating Instructions - CPG2500 Vacuum Calibrator...
  • Page 33 Vacuum Calibrator CPG2500 6.4.3.3 Auxiliary Displays The Transducer Channel Auxiliary Display(s) can be set in the Transducer App by selecting Auxiliary Display 1, 2 or 3 and selecting from, None, Peak, Rate or Units. Figure 6.4.3.3-A shows auxiliary display 1, 2 and 3 set for units, peak and rate respectively. Auxiliary displays will appear in the Measure App as seen in Figure 6.4.3.3-B.
  • Page 34 Vacuum Calibrator CPG2500 6.4.4 Remote Application With the remote application you can select the remote command set for all interfaces. The GPIB address, Ethernet network parameters and Serial parameters can also be set here. Details about the Remote Operation (command sets, cable requirements, etc.) can be found in Section 7, Remote Operation.
  • Page 35 Vacuum Calibrator CPG2500 6.4.4.2 Remote Communication Settings The remainder of the Parameters in the Remote Application present the choice of a numeric entry, a numeric address entry, or a radio button selection. The parameters that require a numeric entry will present a numeric keypad with min and max limits for the variable.
  • Page 36 Vacuum Calibrator CPG2500 6.4.5 Info Application The Info application displays information about the instrument, including: • Mensor address, and email • Model number, serial number and operating software version. • Internal Sensor model number, serial number, range, software version Figure 6.4.5 - Info Application Operating Instructions - CPG2500 Vacuum Calibrator...
  • Page 37 Vacuum Calibrator CPG2500 6.4.6 Leak Test Application The Leak Test application does not function with this model of the instrument. 6.4.7 Troubleshooting Application The Troubleshoot Application will display information about error conditions and remote communications. Within the troubleshoot screen (Figure 6.4.7-B), push the Error button to display any errors that have occurred in the instrument due to a communication or network error.
  • Page 38 Vacuum Calibrator CPG2500 6.4.8 Logging Application The logging application enables real-time storage of internal transducer output and state in a USB device plugged to the front panel of the CPG2500. The application also provides a means to log any remote communication (commands and responses) to the instrument. The primary objective of this application is to provide test data for troubleshooting purposes.
  • Page 39 Vacuum Calibrator CPG2500 6.4.8.2 Pressure Logging The Logging App allows the storage of the time interval, active pressure and rate reading from all channels of the instrument to a file. The “File format” button provides selection of “.DAT” or “.CSV” file type.
  • Page 40 Vacuum Calibrator CPG2500 6.4.9 Service Application The service application is a password protected area where calibration of all connected sensors can be accomplished. In addition, this is where the password for entering this area can be changed. Figure 6.4.9-A - Service Application (locked) Press the Enter button to show the numeric keypad to enter a password.
  • Page 41 Vacuum Calibrator CPG2500 6.4.10 Unlocked Service Application After the Password has been entered, the unlocked Service Application will appear (Figure 6.4.10). To re-lock this screen, press the lock button. Figure 6.4.10 - Unlocked Service Application From the Unlocked Service Application, the Password can be changed by pressing the Enter button next to the Change Password label.
  • Page 42 Vacuum Calibrator CPG2500 Remote Operation Use the screens in Section 6.4.4 Remote Application to set the operating parameters for the instrument command set, Ethernet, Serial (RS-232) and IEEE-488 (GPIB) information. Command Set Command Set button – Users’ can select which model remote protocol they would like to emulate for simulation and testing purposes.
  • Page 43 Vacuum Calibrator CPG2500 Ethernet The Ethernet function allows the user to set the following by inputting a numeric value in each separate field: • IP • Netmask • Gateway • Port • DHCP settings Set the Ethernet communication parameters as described in Section 6.4.5. CAUTION: Please contact your network administrator for proper settings.
  • Page 44 Vacuum Calibrator CPG2500 Serial Set the Serial communication parameters as shown in Section 6.4.5 Remote Application. The serial communication port allows the CPG2500 to communicate in RS-232 format with computers, terminals, PDAs, or similar hosts. These parameters should be set to match your host computer. Default settings are: 57600 baud, 8 data bits, 1 stop bit, no parity, and no echo.
  • Page 45 Vacuum Calibrator CPG2500 Mensor Command Set This Mensor command set is the default on the CPG2500. For queries (ending with a ?), the Data column represents the response of the CPG2500. All response strings begin with a space character or an “E” representing that there is an error in the error queue.
  • Page 46 Vacuum Calibrator CPG2500 Output Formats Pressure readings are returned in exponential notation in a format according to the OUTFORM command as follows. Outform applies to both pressure channels. 1. <sp> pressure value <cr><lf> 2. <sp> pressure, units number,STANDBY <cr><lf> 3. <sp> pressure, pressure rate <cr><lf> 4.
  • Page 47 Vacuum Calibrator CPG2500 Brightness Brightness f Set screen brightness between 0 and 100% Brightness? <sp>n.nnnnnE+nn<cr><lf> Read the screen brightness Asset_tag? <sp>sssssssss<cr><lf> Return customer asset tag string Caldisable YES,NO Sets whether or not the ability to cali- brate the sensors is disabled Caldisable? <sp>(YES or NO)<cr><lf>...
  • Page 48 Vacuum Calibrator CPG2500 Gasdensity? <sp>+n.nnnnnE+nn<cr><lf> Gets the head pressure gas density in lb/cuft. Only set for currently active channel. See alias command MEDIADENSITY? Gastemp Value in degrees Degrees in ∞F or ∞C depending on selected units category. See HEADUNITS? channel. See alias command MEDIATEMP Gastemp? <sp>+n.nnnnnE+nn<cr><lf>...
  • Page 49 Vacuum Calibrator CPG2500 LANGUAGE XXXX Set the displayed language. String to be sent may be ENGLISH, ENG- LISH GB, ENGLISH CA, GERMAN, FRENCH, FRENCH CH, FRENCH CA, SPANISH, SPANISH LA, ITALIAN, POLISH, PORTUGUESE, PORTU- GUESE BR, RUSSIAN, JAPANESE, CHINESE, KOREAN LANGUAGE? <sp>XXXX<cr><lf>...
  • Page 50 Vacuum Calibrator CPG2500 Outform 1 to 4 – see Section 7.9 Sets the output format Outform? <sp>X<cr><lf> Returns the output format – see table below Peakmax? <sp>n.nnnnnE+nn<cr><lf> Returns the maximum pressure since peakreset was sent Peakmin? <sp>n.nnnnnE+nn<cr><lf> Returns the minimum pressure since peakreset was sent Peakreset None...
  • Page 51 Vacuum Calibrator CPG2500 Sparity? <sp>CCCC<cr><lf> Returns the serial parity Sstop 1 or 2 Sets the serial stop bits Sstop? <sp>X<cr><lf> Returns the serial stop bits SUBUNITS? <sp>XXXX<cr><lf> Returns the instrument auxiliary dis- play units in a text string Transfer_factory_to_lin- Copy factory linearity coefficients to earity customer Units...
  • Page 52 Vacuum Calibrator CPG2500 7.9.1 Units Command Syntax for Measurement Units Description Output Format Type pounds per square inch Imperial inches of mercury @ 0°C INHG Imperial inches of mercury @ 60°F INHG Imperial inches of water @ 4°C INH2O Imperial inches of water @ 20°C INH2O Imperial...
  • Page 53 Vacuum Calibrator CPG2500 7.9.2 CPG2500 Error Codes Code Serial Poll Byte Description Error String Returned No errors NO ERRORS Parameter error EGPIB PARAMETER ERROR: String that was sent Syntax error EGPIB SYNTAX ERROR: String that was sent 7.9.3 GPIB Capability Codes Full source handshake capability Full acceptor handshake capability Talker with serial poll and unaddress if MLA...
  • Page 54 Vacuum Calibrator CPG2500 Accessories • Remote Transducer Cable • Vacuum Pump Accessory Kit External Pressure Sensor Connection Cable A shielded ribbon cable is required to connect an external transducer (DUT) and can be supplied. The CPT6100/6180 part number is: 4060090001. Vacuum Pump Accessory Kit A vacuum pump accessory kit can be ordered as an option to the CPG2500 Vacuum Calibrator to ensure optimal vacuum source for the internal transducer and the DUT.
  • Page 55 Vacuum Calibrator CPG2500 Maintenance The CPG2500 was designed for maintenance-free operation. User maintenance is not recommended. If you have questions not covered by this manual, call 1.800.984.4200 (USA only), or 1.512.396.4200 for assistance, or send an e-mail to tech.support@mensor.com. Beyond the Warranty Take advantage of Mensor’s expert product care.
  • Page 56 Vacuum Calibrator CPG2500 10 Calibration The CPG2500 automatically adjusts the pressure reading for the effects of temperature and non-linearity within the calibrated temperature range of 15-40°C. The process is referred to as dynamic compensation because each reading is so adjusted before it is output to the display or to a communication bus. Thus, a calibrated CPG2500 operated within its temperature band, and with proper zero and span adjustments, will provide accurate pressure measurements.
  • Page 57 Vacuum Calibrator CPG2500 10.4 Service Application The Service application is a password protected area where calibration of all connected sensors can be accomplished. In addition, this is where the password for entering this area can be changed. Figure 10.4-A - Service Application (locked) Press the Enter button to show the numeric keypad to enter a password.
  • Page 58 Vacuum Calibrator CPG2500 10.4.1 Service Application (unlocked) After the Password has been entered, the unlocked Service Application will appear (Figure 10.4.1). To re-lock this screen, press the Lock button. Figure 10.4.1 - Unlocked Service Application From the unlocked Service Application, the Password can be changed by pressing the Enter button next to the Change Password label.
  • Page 59 Vacuum Calibrator CPG2500 10.5 Calibration Data The Calibration Data Application is where the calibration data for each transducer is stored and amended. The Serial number (S/N), Zero offset (Zero) and Span offset (Span) can be seen in this screen. The date of calibration, the calibration interval and the certificate number can be entered by pressing the corresponding button, then saved by pressing the Check [ ] button.
  • Page 60 Vacuum Calibrator CPG2500 Figure 10.7 - Two Point Cal Application 10.8 Linearization The Linearization Application is not applicable to this instrument. Operating Instructions - CPG2500 Vacuum Calibrator...
  • Page 61 Vacuum Calibrator CPG2500 10.9 Head Pressure The Head Pressure Application provides an automated way to calculate the head pressure offset between a device being tested and the CPG2500 sensor, based on: • Local Gravity • Media: pneumatic • Height: the difference between the device under test and the sensing element in the transducers (internal and external) in the CPG2500.
  • Page 62 Vacuum Calibrator CPG2500 11 Appendix Table 11.1 - Measurement Units (unitno) Code Description Output Format pounds per square inch inches of mercury @ 0°C inHg 0°C inches of mercury @ 60°F inHg 60°F inches of water @ 4°C inH2O 4°C inches of water @ 20°C inH2O 20°C inches of water @ 60°F...
  • Page 63 Vacuum Calibrator CPG2500 User Units 1 User defined User Units 2 User defined 11.1 Conversion Factors, PSI The values listed in the column “To convert from PSI” are the values imbedded in the instrument program. The values listed under “To convert to PSI” are internally calculated approximations based on the imbedded values.
  • Page 64 Vacuum Calibrator CPG2500 0.006894757 145.0377 mmH2O 20°C 704.336 0.001419777 cmH2O 20°C 70.4336 0.01419777 MH2O 20°C 0.704336 1.419777 11.2 Conversion Factors, Millitorr The following table lists factors which should be used as multipliers when converting other pressure units to or from millitorr. Table 11.3 - Conversion Factors, millitorr Code Pressure Unit...
  • Page 65 Vacuum Calibrator CPG2500 0.0000001333220 7500636.259 mmH2O 20°C 0.01361955 73.42388114 cmH2O 20°C 0.001361955 734.2388114 MH2O 20°C 0.00001361955 73423.88114 Operating Instructions - CPG2500 Vacuum Calibrator...
  • Page 66 Vacuum Calibrator CPG2500 11.3 Conversion Factors, Pascal The following table lists factors which should be used as multipliers when converting other pressure units to or from Pascal. 11.4 - Conversion Factors, Pascal Unit No. Pressure Unit To convert from Pascal To convert to Pascal 1.450377E-04 6.894757E+03...
  • Page 67 Tel: 512-396-4200 Website: www.mensor.com Fax: 512-396-1820 Email: sales@mensor.com Vacuum Calibrator CPG2500 PN 0018906001 • 10/2019 WIKA Alexander Wiegand SE & Co. KG Alexander-Wiegand-Straße 30 D-63911 Klingenberg / Germany Tel: (+49) 93 72/132-5015 Website: www.wika.de Fax: (+49) 93 72/132-8767 Email: CTsales@wika.de...