(this can take up to 24 hours). Once you have secured an account, use the Teledyne Marine software portal to access this data with your unique username and password.
Page 4
OTES Page iv EAR99 Technology Subject to Restrictions Contained on the Cover Page.
P/N 95B-6149-00 (JULY 2023) System Turnaround The Pinnacle 45 kHz Self-Contained ADCP allows for turn-around time of a mooring deployment well un- der 60-minutes. During a typical turn, the data needs to be transferred, the batteries replaced, and the system prepared to redeploy.
P/N 95B-6149-00 (JULY 2023) Pinnacle 45 Self-Contained Deployment Guide Preparing the Pinnacle ADCP R E P A R I N G T H E I N N A C L E I N C L U D E S T H E F O L L O W I N G S T E P S...
Page 7
Pinnacle 45 Self-Contained Deployment Guide P/N 95B-6149-00 (JULY 2023) Place the end-cap on the housing, aligning the notches embossed on the end-cap with the notch embossed on the housing. Examine the end-cap assembly captured bolts and inserts for corrosion; replace if necessary.
P/N 95B-6149-00 (JULY 2023) Pinnacle 45 Self-Contained Deployment Guide Connecting to the Pinnacle O N N E C T I N G T O T H E I N N A C L E I N C L U D E S T H E F O L L O W I N G S T E P S...
Page 9
Pinnacle 45 Self-Contained Deployment Guide P/N 95B-6149-00 (JULY 2023) LED Indications There is one blue LED mounted on the side of the transducer housing. The blink rates indicate the Pinna- cle status as defined below. The LED blinks in an 8-bit pattern representing two seconds of time. Each bit represents 0.25 seconds.
Page 10
P/N 95B-6149-00 (JULY 2023) Pinnacle 45 Self-Contained Deployment Guide Quick Review For help on using Pinnacle Utilities, click icon. Always close the TRDI software connection to the system by clicking on the X ( ) in the Connect to the Pinnacle session tab and then power down the Pinnacle before disconnecting the Ethernet cable.
Pinnacle 45 Self-Contained Deployment Guide P/N 95B-6149-00 (JULY 2023) Planning the Deployment L A N N I N G T H E E P L O Y M E N T I N C L U D E S T H E F O L L O W I N G S T E P S...
Page 12
P/N 95B-6149-00 (JULY 2023) Pinnacle 45 Self-Contained Deployment Guide Set the Schedule tab as needed. This is the default setting when the Bursts Wizard is selected. Click the Change button on the Output tab. Use this screen to set up the Input and Output triggers if needed.
Pinnacle 45 Self-Contained Deployment Guide P/N 95B-6149-00 (JULY 2023) Deploying the Pinnacle E P L O Y I N G T H E I N N A C L E I N C L U D E S T H E F O L L O W I N G S T E P S...
Page 14
P/N 95B-6149-00 (JULY 2023) Pinnacle 45 Self-Contained Deployment Guide Testing the Pinnacle To run the Built-in tests: Setup the communication parameters between Pinnacle Utilities and the Pinnacle. Place the Pinnacle system in enough water to cover the transducer face. Use wood strips or a hoist to lift the transducer to protect the face.
Page 15
Pinnacle 45 Self-Contained Deployment Guide P/N 95B-6149-00 (JULY 2023) Testing the Sensors To view the sensor data: Start Pinnacle Utilities and connect to the Pinnacle ADCP. Click Open on the System + Sensors box. Click Start on the Sensor Data button.
Page 16
P/N 95B-6149-00 (JULY 2023) Pinnacle 45 Self-Contained Deployment Guide Testing for Acoustic Interference Observation of the acoustic interference plots will provide information about interference to the Pinnacle ADCP. The Pinnacle Utilities software provides the user with a frequency domain plot (~100% band- width) of the Pinnacle’s four receiver channels during a sampling interval.
Page 17
Pinnacle 45 Self-Contained Deployment Guide P/N 95B-6149-00 (JULY 2023) In an oil & gas environment, it is typically impossible to meet the first two requirements above. Oil fields are full of metal structures that preclude a “magnetically clean” environment. Even in standard oceanography, for example a deep-water deployment in the...
Page 18
P/N 95B-6149-00 (JULY 2023) Pinnacle 45 Self-Contained Deployment Guide Deploying the Pinnacle To start a Self-Contained deployment: Start Pinnacle Utilities and connect to the Pinnacle ADCP. Run the pre-deployment checks. On the System + Sensors screen, change the System Type to SC.
Page 19
Pinnacle 45 Self-Contained Deployment Guide P/N 95B-6149-00 (JULY 2023) Remove the cable and install dummy plugs on both connectors. Thread the coupling ring onto the receptacle to complete the connection. The Pinnacle ADCP is now ready to deploy. Quick Review...
P/N 95B-6149-00 (JULY 2023) Pinnacle 45 Self-Contained Deployment Guide Recovering Data E C O V E R I N G A T A I N C L U D E S T H E F O L L O W I N G S T E P S...
Page 21
Pinnacle 45 Self-Contained Deployment Guide P/N 95B-6149-00 (JULY 2023) Downloading Data Files Upon completion of stopping the deployment and verifying that the removable recorder and fixed re- corder match, data can be transferred or downloaded to a computer. To transfer data: Remove the purge plug.
Page 22
P/N 95B-6149-00 (JULY 2023) Pinnacle 45 Self-Contained Deployment Guide Erasing the Recorder Once all data has been transferred or downloaded, the recorders can be erased and formatted for the next deployment. To format the recorders: Start Pinnacle Utilities and connect to the Pinnacle ADCP via Ethernet.
Page 23
Pinnacle 45 Self-Contained Deployment Guide P/N 95B-6149-00 (JULY 2023) Viewing Data Pinnacle Utilities can open and display Pinnacle data files. To open a data file: Start Pinnacle Utilities. On the start screen, recently opened deployments are shown in the box and can be clicked to reopen the file.
Page 24
P/N 95B-6149-00 (JULY 2023) Pinnacle 45 Self-Contained Deployment Guide Checking the No button on the Use Correction box allows the user to operate on their original raw data. When done viewing the post-calibration data, click Close and then the data file will open.
Use this appendix if you are a LongRanger user and now using a Pinnacle system PDDecoder Library in C language The Teledyne Marine PDDecoder library is an open-source library written in C language to decode the PD0 data formats that are commonly output by Teledyne Marine/Teledyne RD Instruments ADCPs.
Page 26
P/N 95B-6149-00 (JULY 2023) Pinnacle 45 Self-Contained Deployment Guide OTES Page 22 EAR99 Technology Subject to Restrictions Contained on the Cover Page.