Technical data
9.1 Functional data
Variable
Degree of protection according to EN 60 529
Maximum tightening torque of the fixing screws
Cable connection
9.1.4

SIWAREX WL260 SP-S AE

Table 9-4
Variable
Type series
Designation
Construction type
Possible applications
Rated load E
Accuracy class
Combined error F
Deviation F
Creep error F
30 min
Temperature coefficient
Characteristic value T
Zero signal T
Min. initial loading E
Maximum working load L
Break load L
Maximum lateral load L
Rated measuring path h
For E
For E
Recommended supply voltage
Rated characteristic value C
Tolerance of zero signal D
Input resistance R
Output resistance R
Insulation resistance R
Rated temperature range B
Operating temperature range B
Storage temperature range B
Sensor material
Degree of protection according to EN 60 529
52
Technical data SIWAREX WL260 SP-S AE
max
comb
V
cr
Kc
Ko
min
u
d
lq
n
= 0.3; 0.6 kg:
max
= 1.2; 1.5; 3 kg:
max
Rated
o
i
o
is
tn
ts
t
Operating Instructions, 04/2021, A5E02199611B-12
Value
IP67
166 Nm
Six-core, shielded
PVC cable 3 m
Value
WL260
SP-S AE
Single-point load cell
Small platform scales and small conveyor
scales
0.3; 0.6; 1; 1.2; 1.5; 3.0 kg
0.015%
± 0.015% C
n
± 0.017% C
n
≤ ± 0.015% C
n
≤ ±0.03% C
/ 10 K
n
≤ ±0.03% C
/ 10 K
n
0 kg
120% E
max
300% E
max
250% E
max
0.25 mm
0.22 mm
5 ... 12 V DC
0.9 ± 0.1 mV/V
< ± 2.0% C
n
406 ± 6 Ω
350 ± 3 Ω
5000 MΩ at 50 V DC
-10 to +40 °C
-20 to +50 °C
-20 to +20 °C
Aluminum
IP65
SIWAREX WL200