Cochlear Nucleus 8 Manual page 25

Hybrid mode
Hide thumbs Also See for Nucleus 8:

Advertisement

Available languages

Available languages

Problem
Possible cause
Painful fit
Speaker unit is
incorrectly inserted
into the ear
Hybrid system is not
properly fitted
Wax buildup in the
outer ear canal
Output lower
Microphone cover
than expected
needs changing
Loose Hybrid
receiver connection
Incorrect program
Hybrid receiver is
broken
Hybrid Mode Professionals Guide
Test / resolution
Insert correctly and re-train the recipient in correct
insertion.
1. Check areas of the outer ear where the sound
processor sits for sore spots or friction marks.
2. Check the Hybrid receiver cable length. If it is too
long, the cable will not sit nicely against the ear
and the speaker unit may sit too low in the ear
canal. If it is too short, the dome may keep
falling out. Adjust so it sits snugly and without
tension. See Shape the Hybrid receiver cable on
page 22.
Check for sore spots using an otoscope – these are
usually caused by wax buildup. Refer the patient to a
medical practitioner to have their ear canal cleaned.
Change microphone cover.
Check the Hybrid receiver connection to the earhook
and replace any damaged parts.
Perform a listening test. Use Acoustic Test Box or
Hearing Instrument Test Box if available and cross
check against target curve in the Custom Sound
software.
Check the Hybrid receiver using a listening test, and
replace it if required.
If the test is not conclusive, reproduce the OSPL90
curve from the Data Sheet and check against the
outlined limits.
21