11 Error messages

Error
Description
message
Err lo
Too small calibration weight
Err hi
Too big calibration weight
Err 3
Calibration error
Err 4
Resetting range error
Err 6
Value of the analog-to-digital
converter
Err 7
Error of percentage weighing
Err 19
Zero point drift
Err E
EPROM memory error
--------------
Maximum load exceeded
38
Possible causes /
Remedy error
Check calibration weight and calibrate
again
Resetting range exceeding when
switched on
Reduce load of the weighing scale
plate
Reduce load of the weighing scale
plate
Check if the weighing scale plate is
placed correctly
Check if the load cell is connected
correctly
Reference weight when determining
percentage weight must be > 0.5 d
Remove additional preloads (vessels).
Adjustment of weighing scale
Check EPROM memory
Reduce the weight or reduce load of
the weighing scale plate
EWJ-BA-e-1917