Specification
1
Product Name:
Size: 83.5 mm (L) × 54 mm (W) × 19 mm (Thickness)
2
Glucose Test Range: 1.1-33.3 mmol/L (20-600 mg/dL)
3
Results Display: plasma equivalent
3
Minimum Sample Volume: 0.5µL
4
Test Time: 5 seconds
6
Battery: CR 2032 3.0V coin cell battery
7
Battery life: >1,000 readings(wireless functions disabled)
8
Glucose Concentration Units: mmol/L or mg/dL depending on the
8
standard of your country
8
Memory Storage: 500 test results with date and time stamp
Auto Shutdown: Automatic shutdown after 2 minutes
Display Size: 40mm × 42mm
Weight: about 50 grams (including battery)
Operating Temperature: 5-45°C (41-113°F)
Operating Humidity: 20-90% (non-condensing)
Hematocrit Range: 10-70%
Note:
Blood containing high levels of vitamin C or other reducing
agents can lead to inaccurate test results. Triglycerides
above 3,000 mg/dL and cholesterol above 500 mg/dL will lead to
inaccurate test results.
Blood Glucose Monitoring System